Agilent-technologies Multiple Channel Controller 85330A Manuale Utente Pagina 34

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 33
2-4 85330A Multiple Channel Controller
Performance Verification
Performance Verification Test
Digital Oscilloscope Set up the Agilent 54503A digital oscilloscope (or any oscilloscope) as
described in the following procedure, using the same settings.
NOTE The oscilloscope setup described in the following procedure will not display
a trace on the oscilloscope screen until a pulse is initiated. The oscilloscope
will then hold the trace until the [CLEAR DISPLAY] is pressed.
Procedure
Turn ON the oscilloscope and press the following keys on the 54503A:
[CHAN]: {1}
{ON}
{2v/Div}
{Offset: 0v}
{Coupling: DC}
{Input: 1 M
}
{Probe Ratio: 1:1}
[CHAN]: {2}
{ON}}
{2v/Div}
{Offset: 5v}
{Coupling: DC}
{Input: 1 M
}
{Probe Ratio: 1:1}
[TRIG]:
{Trig’d}
{Edge}
{Source: 1}
{Adjust: 1v}
{ } (- Edge)
[TIMEBASE]:
{2 us/Div}
{Delay: 0}
{Reference: Cntr}
[SAVE] [1]
Vedere la pagina 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 165 166

Commenti su questo manuale

Nessun commento