2-16 85330A Multiple Channel Controller
Performance Verification
Verifying Remote Ports 1 and 2
Remote Port 1 and remote Port 2 Test Procedure
1. Download the 85330A drivers version A.03.00 or above.
2. When
select instrument is displayed, press:
{85330A}, {More}, {TEST}, {More}
3. Connect the modified connector to REMOTE 1 on the rear panel and
press:
{REM_1_2}
4. The display should read:
“85330A_144:REM1 0101 PASS, REM2 1100 FAIL”
5. Move the modified connector to REMOTE 2 on the rear panel and
press:
{REM_1_2}
6. The display should read:
“85330A_144: REM1 1111 FAIL, REM2 0011 PASS”