7-12 85330A Multiple Channel Controller
Remote Programming Command Reference
Standard Command Reference
RUNTime:SWITch:DELay
Used for the delay between asserting the switch address list entry of the
RUNT:SWIT:SCAN command and the issuance of a receiver trigger. The
input to the command is in microseconds.
Default The default is 2 microseconds.
Syntax RUNT:SWIT:DEL <microseconds>;
Example RUNT:SWIT:DEL 50; ! Set the trigger or setting
! time to 50 micro-seconds.
NOTE When using the multiple-parameter display mode of the 8530A, the
minimum switch settling time setting is 50
µs. Using faster settling times
can cause problems during the measurement.
For the Fast Data Acquisition modes, 2 µs settling time is recommended.