Agilent Technologies 22D Guida Utente Pagina 216

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 276
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 215
5-46
Making Measurements
Making Delay and Phase Measurements
Phase
Phase is the calculated phase shift from source 1 to source 2, expressed in
degrees. Negative phase shift values indicate that the rising edge of source 1
occurred after the rising edge of source 2. Press the Settings softkey to select
the source channels for the phase measurement.
•Press the Settings softkey to select the source 1 and source 2 channels
for the phase measurement.
The default Phase settings measure from channel 1 to channel 2.
The example below shows a phase measurement between the channel 1 and
the math d/dt function on channel 1.
Phase Measurement
Phase =
Delay
Source 1 Period
--------------------------------------------
X 360
Source 1
Source 2
Delay
Period
Vedere la pagina 215
1 2 ... 211 212 213 214 215 216 217 218 219 220 221 ... 275 276

Commenti su questo manuale

Nessun commento