Making Measurements
Making overshoot and preshoot measurements
Overshoot
Overshoot is distortion that follows a major edge transition expressed as a
percentage of Amplitude. The X cursors show which edge is being measured
(edge closest to the trigger reference point).
Automatic Overshoot measurement
Rising edge overshoot =
local Maximum – Top
Amplitude
-------------------------------------------------------
X 100
Falling edge overshoot =
Base – local Minimum
Amplitude
--------------------------------------------------------
X 100
local Maximum
Overshoot
Overshoot
local Minimum
Top
Base