2-18
Making Mixer Measurements (Option 089 Only)
Conversion Loss Using the Frequency Offset Mode
Performing the R-Channel Measurement
1. Connect the equipment as shown in Figure 2-15.
Figure 2-15 R-Channel Mixer Measurement Equipment Setup
The analyzer is now displaying the conversion loss of the mixer calibrated with power
meter accuracy.
2. To view the conversion loss in the best vertical resolution, press
.
Scale Ref
AUTOSCALE
Vedere la pagina 148