Agilent Technologies 24A Guida Utente Pagina 191

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 453
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 190
3-11
Making Time Domain Measurements
Making Reflection Response Measurements
8. To position the marker on the reflection of interest, press:
and turn the front panel knob, or enter a value from the front panel keypad.
In this example, the velocity factor was set to one-half the actual value, so the marker
reads the time and distance to the reflection.
9. To position a marker at each reflection of interest, as shown in Figure 3-9, press:
, turning the front panel knob or entering a
value from the front panel keypad after each key press.
Figure 3-9 Device Response in the Time Domain
Marker
MARKER 2
MARKER 3
MARKER 4
Vedere la pagina 190