Agilent Technologies 24A Guida Utente Pagina 90

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 453
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 89
1-77
Making Measurements
Using Limit Lines to Test a Device
2. To establish the start frequency and limits for a sloping limit line that tests the low side
of the filter, press:
3. To terminate the lines and create a sloping limit line, press:
4. To establish the start frequency and limits for a sloping limit line that tests the high
side of the filter, press:
5. To terminate the lines and create a sloping limit line, press:
You could use this type of limit to test the shape factor of a filter.
Figure 1-61 Sloping Limit Lines
ADD
STIMULUS VALUE
123 M/µ
UPPER LIMIT
65 x1
LOWER LIMIT
200 x1
DONE
LIMIT TYPE
SLOPING LINE
RETURN
ADD
STIMULUS VALUE
125 M/µ
UPPER LIMIT
26 x1
LOWER LIMIT
200
x1
DONE
LIMIT TYPE
SINGLE POINT
RETURN
ADD
STIMULUS VALUE
144 M/µ
UPPER LIMIT
26 x1
LOWER LIMIT
200 x1
DONE
LIMIT TYPE
SLOPING LINE
RETURN
ADD
STIMULUS VALUE
146 M/µ
UPPER LIMIT
65 x1
LOWER LIMIT
200 x1
DONE
LIMIT TYPE
SINGLE POINT
RETURN
Vedere la pagina 89
1 2 ... 85 86 87 88 89 90 91 92 93 94 95 ... 452 453

Commenti su questo manuale

Nessun commento