Module 6
Low Current Measurement
6-12
Do not change this screen
ZERO CHECK
Measurement Setup
This default sweep setup is enough for the Zero Check measurement.
The sweep steps are small (10 mV) and that is ideal for checking current at fA levels.
Also the sweep starts at 0, so no big discontinuities on the first step of the sweep.
There is no need to change anything on the Measurement Setup screen. But the ranging mode and
the integration time settings must be changed. See the next page.
Click Range button to change the ranging mode.
Click ADC/Integ button to change the integration time setting.