Agilent B1500A User’s Guide, Edition 13 3- 29
Installation
Mounting Connectors
Mounting Connectors
Previous sections described how to install the available accessories, and Agilent
16495 Installation Guide provides the information on how to install connector
plates on the shielding box. However, you may choose to mount connectors directly
on your own connector plate or test fixture.
To do this, you will need to make connector holes and mount the connectors
yourself. This section contains information on connector hole dimensions.
To mount connectors:
1. Select the appropriate parts for your application. See Table 3-5.
2. Make the holes and mount the connectors. See Table 3-6.
For Kelvin connections (remote sensing), use Agilent 16493K Kelvin triaxial
cable. The Kelvin triaxial cable requires a Kelvin triaxial connector, which has
two connector holes and three screw holes.
3. Build the interlock circuit shown in “To Make an Interlock Circuit” on page
3-31.
4. Attach cables from the connectors to the DUT (device under test). See “To
Connect GNDU Output” on page 3-34, “To Connect SMU Output” on page
3-35, and “To Connect MFCMU Output” on page 3-42.