Specifications
Hardware
Table 4-1 Voltage range, resolution, and accuracy (high resolution ADC)
Table 4-2 Current range, resolution, and accuracy (high resolution ADC)
Range
Force
resolution
Measurement
resolution
Force accuracy
±(%+mV)
a
a. ±(% of output/measured value + offset value in mV)
Measurement accuracy
±(%+mV)
a
Maximum
current
±0.5 V 25 V 0.5 V ±(0.018 + 0.15) ±(0.01 + 0.12) 100 mA
±2 V 100 V 2 V ±(0.018 + 0.4) ±(0.01 + 0.14) 100 mA
±5 V 250 V 5 V ±(0.018 + 0.75) ±(0.009 + 0.25) 100 mA
±20 V 1 mV 20 V ±(0.018 + 3) ±(0.009 + 0.9) 100 mA
±40 V 2 mV 40 V ±(0.018 + 6) ±(0.01 + 1)
b
b. 100 mA (Vo 20 V), 50 mA (20 V < Vo 40 V), 20 mA (40 V < Vo), Vo is the output voltage.
±100 V 5 mV 100 V ±(0.018 + 15) ±(0.012 + 2.5)
Range
Force
resolution
Measurement
resolution
Force accuracy ±(%+A+A)
a
a. ±(% of output/measured value + fixed offset in A + proportional offset in A), Vo is the output voltage in V.
Measurement accuracy
±(%+A+A)
a
Maximum
voltage
±1 pA
b
b. Applicable if Agilent E5288A Atto Sense and Switch Unit (ASU) is used.
1 fA 0.1 fA ±(0.9+15E-15) ±(0.9+12E-15) 100 V
±10 pA
c
c. Not applicable for the MPSMU without ASU.
5 fA
0.4 fA
b
/ 1 fA
±(0.46+3E-14+VoE-17) ±(0.46+15E-15+Vo1E-17) 100 V
±100 pA
c
5 fA
0.5 fA
b
/ 2 fA
±(0.3+1E-13+VoE-16) ±(0.3+3E-14+Vo1E-16) 100 V
±1 nA 50 fA 10 fA ±(0.1+3E-13+Vo1E-15) ±(0.1+2E-13+Vo1E-15) 100 V
±10 nA 500 fA 10 fA ±(0.1+3E-12+Vo1E-14) ±(0.1+1E-12+Vo1E-14) 100 V
±100 nA 5 pA 100 fA ±(0.05+3E-11+Vo1E-13) ±(0.05+2E-11+Vo1E-13) 100 V
±1 A 50 pA 1 pA ±(0.05+3E-10+Vo1E-12) ±(0.05+1E-10+Vo1E-12) 100 V
±10 A 500 pA 10 pA ±(0.05+3E-9+Vo1E-11) ±(0.04+2E-9+Vo1E-11) 100 V
±100 A 5 nA 100 pA ±(0.035+15E-9+Vo1E-10) ±(0.03+3E-9+Vo1E-10) 100 V
±1 mA 50 nA 1 nA ±(0.04+15E-8+Vo1E-9) ±(0.03+6E-8+Vo1E-9) 100 V
±10 mA 500 nA 10 nA ±(0.04+15E-7+Vo1E-8) ±(0.03+2E-7+Vo1E-8) 100 V
±100 mA 5 A 100 nA ±(0.045+15E-6+Vo1E-7) ±(0.04+6E-6+Vo1E-7)
d
d. 100 V (Io 20 mA), 40 V (20 mA < Io 50 mA), 20 V (50 mA < Io), Io is the output current.
Vedere la pagina 148