Agilent Technologies B1500A Guida Utente Pagina 62

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 202
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 61
2- 20 Agilent B1500A User’s Guide, Edition 13
Introduction
Measurement Modules
Table 2-1 HPSMU Output Voltage Ranges and Resolutions
Table 2-2 HPSMU Output Current Ranges and Resolutions
Range Output Value
Setting
Resolution
Maximum
Current
2 V 0  |V| 2 V 100 V ± 1000 mA
20 V 0  |V| 20 V 1 mV ± 1000 mA
40 V 0  |V| 40 V 2 mV ± 500 mA
100 V 0  |V| 100 V 5 mV ± 125 mA
200 V 0  |V| 200 V 10 mV ± 50 mA
Range Output Value
Setting
Resolution
Maximum
Voltage
1 nA 0  |I| 1.15 nA 50 fA ± 200 V
10 nA 0  |I| 11.5 nA 500 fA ± 200 V
100 nA 0  |I| 115 nA 5 pA ± 200 V
1 A 0  |I| 1.15 A 50 pA ± 200 V
10 A 0  |I| 11.5 A 500 pA ± 200 V
100 A 0  |I| 115 A 5 nA ± 200 V
1 mA 0  |I| 1.15 mA 50 nA ± 200 V
10 mA 0  |I| 11.5 mA 500 nA ± 200 V
100 mA 0  |I| 50 mA 5 A ± 200 V
50 mA < |I| 115 mA 5 A ± 100 V
1 A 0  |I| 50 mA 50 A ± 200 V
50 mA < |I| 125 mA 50 A ± 100 V
125 mA < |I| 500 mA 50 A ± 40 V
500 mA < |I| 1 A 50 A ± 20 V
Vedere la pagina 61
1 2 ... 57 58 59 60 61 62 63 64 65 66 67 ... 201 202

Commenti su questo manuale