General Specifications
Environment
For use in indoor facilities
Temperature Range
Operating: 0 °C to +55 °C
Storage: -30 °C to +70 °C
Humidity Range
Operating: 30 % to 80 % RH, non-condensing
Storage: 10 % to 90 % RH, non-condensing
Altitude
Operating: 0 m to 2000 m
Storage: 0 m to 4600 m
Power Supply
AC voltage: 90 V to 264 V
Line frequency: 47 Hz to 63 Hz
Maximum volt-amps: 250 VA
Cooling
Forced air. Side intake and rear exhaust.
EMC
IEC61326-1/EN61326-1
AS/NZS CISPR 11
KC: RRA Notification amending Radio Waves Act Article 58-2
Safety
IEC61010-1/EN61010-1
CAN/CSA-C22.2 No. 61010-1-04, C/US
Certifications
CE, cCSAus, C-Tick, KC
Dimensions
Without handle & feet: 88 mm H x 213 mm W x 450 mm D
Bench Configuration (with handle & feet): 180 mm H x 260 mm W x 480 mm D
Weight
B2901A/B2911A: 5.0 kg
B2902A/B2912A: 6.4 kg
Operation and Functions
Front Panel Interface
4.3” Color TFT WQVGA (480x272, with LED backlight) provides a graphical user
interface (GUI) with the following features.
– 4-View Modes (Single, Dual, Graph, Roll view)
– Easy-to-use key operations
– Easy access to basic and advanced source/measurement capabilities
Single view
Single view provides basic and advanced settings and display capabilities for the
selected channel from the front panel of the instrument. No additional controller
or software is required.
Dual view
Dual view provides the basic settings and display capabilities for both channels
1 and 2. Up to 6½ digits can be displayed in dual view mode. This mode is avail-
able only for the B2902A/B2912A.