5
Source Supplemental Characteristics
Temperature coefficient (0 °C to 18 °C and 28 °C to 50 °C):
±(0.1 x Accuracy specification) /°C.
Max output power and source/sink limits:
31.8 W
±6 V@ ±3.03 A, ±21 V@ ±1.515 A, ±210 V@ ±105 mA, four quadrant
source or sink operation.
Current limit/compliance:
Accuracy is same as current source. Minimum value is 1 % of range, or 1 nA
in 10 nA range.
Voltage limit/compliance:
Accuracy is same as voltage source. Minimum value is 1 % of range, or
20 mV in 200 mV range
Over range :
101 % of source range for 1.5 A and 3 A ranges. 105 % of source range other
than 1.5 A and 3 A ranges. No over range for 200 V range with current
exceeding 105 mA pulse only condition.
Over temperature protection:
Output turns off then resets at over temperature sensed internally.
Voltage output settling time:
Time required to reach within 0.1 % of final value at open load condition.
Step is 10 % to 90 % range
200 mV, 2 V Ranges: <50 μs
20 V Range: <110 μs
200 V Range:<700 μs
Slew rate:
≤ 0.36 V / µs, 20 V and 10 mA ranges, 10 MΩ load resistance
Current output settling time:
Time required to reach within 0.1 % (0.3 % for 3 A range)
of final value at short condition. Step is 10 % to 90 % range
10 nA, 100 nA Ranges: <10 ms
1 μA Range: <500 μs
10 μA, 100μA Ranges:<250 μs
1mA to 3 A Ranges:<80 μs
Noise 10 Hz to 20 MHz (V source): 3 mVrms, 20 V range.
V source overshoot:
< ±(0.1 %+10 mV). Step is 10 % to 90 % range, resistive load.
I source overshoot:
< ±0.1 % (<± 0.3 % for 3 A range). Step is 10 % to 90 % range, resistive load.
Voltage source range change overshoot:
≤ 250 mV. 100 kΩ load, 20 MHz bandwidth
Current source range change overshoot:
≤ 250 mV/R load , 20 MHz bandwidth