Low Noise Filter (N1294A-022)
Maximum output range: 210 V / 3 A (DC)
Output connector/connection: Banana, 2-wire/4-wire
Output / residual resistance: 0.3 Ω nominal (2-wire)
Small signal bandwidth: 2 kHz nominal (1 A and 3 A ranges), 800 Hz nominal
(100 mA range)
Noise
Voltage source
0.1 to 10 Hz: Same as voltage specification
10 to 20 MHz : 350 µVrms (20 V / 1.5 A range, 50 Ω load)
Current source
0.1 to 10 Hz: Same as current specification
10 to 1 MHz : 450 µArms (20 V / 1.5 A range, 0.67 Ω load)
Settling time
Voltage source
1
: 640 µs (2-wire/4-wire)
Current source
2
: 1.2 ms (2-wire/4-wire)
1. Time required to reach within 0.1 % of final value at open load condition. Step is 10 % to
90 % range. At 20 V range, 1.5 A limit/1.5 A range
2. Time required to reach within 0.1 % of final value at short condition. Step is 10 % to 90 %
range. At 1.5 A range, 20 V limit / 20V range
Load transient recovery time
Time to recover to within the settling band following a load change.
800 mA current change
1
Settling band ± 20 mV
Time 450 µs (2-wire), 650 µs (4-wire)
1. At a load change to change the flowing current from +100 mA to +900 mA, 10 V voltage force,
20 V range, 1.5 A limit.
Supported ranges:
Voltage: 200 mV to 200 V ranges (210 V maximum)
Current: 10 µA to 3 A ranges (3 A maximum)
Maximum capacitive load: 1 mF
Dimensions: 41.5 mm H x 58.2 mm W x 127.5 mm D (When the filter is inserted
to the output connector of B2961A/B2962A, the depth is 112.5 mm.)
Weight: 0.25 kg
Note:
* The 10 A pulse range and programmable output resistance capability are not
supported by external filters.
* For the current output, only DC is supported when using with external filters.
*The current measurement data monitored by the built-in voltage/current moni
tor can be influenced by the charge and discharge current of the capacitance
inside of the filters.