2-38
Making Measurements
If You Have Problems with a Measurement
Solution 4 - Amplitude Envelope
Check if the Pulse Amplitude is ± 1 dB of the expected value during the useful
part of the burst.
If FM Errors:
Perform a Phase Freq:Data Bits measurement. Dashes (a) will display
under the bits where the power is too low.
If No FM Errors:
Perform pulse demodulation measurements. Connect PULSE
(DEMODULATION OUT) connector to SCOPE IN (MEASURE)
connector on the front panel.
---------