Verifying Performance
Agilent Technologies 8922M/S Specifications
Peak/Transmitter Carrier Power Measurement
RF In/Out only. After one timeslot, 577 µs, from an isolated Receiver Hop Trigger
in the GSM bands.
Power Measurement Accuracy
Pulse ON/OFF Ratio Measurement (Requires Option 006)
ON power is averaged over the useful part of the burst. OFF is averaged over a one
bit interval centered at a user specified time. Non-hopped mode only.
Range -5 to +41 dBm
Input Frequency Setting Error
±10 kHz
Input Level Setting Error
±3 dB
Accuracy
±0.6 dB ± noise effects (+4 to +41 dBm) (0.2 mW)
Supplemental Characteristics:
Accuracy
±0.6 dB ± noise effects (-5 to +4 dBm) (0.2 mW)
Minimum Resolution
0.2 dB
-5 0 5 10 15 20 25 30 35
0.1
0.2
0.3
0.4
0.5
0
Accuracy (dB)
Power (dBm)
Input Frequency Setting Error
±10 kHz
Input Level Setting Error
±3 dB
Timing Accuracy
±1.7 µs (± 1.1 µs typical)
Vedere la pagina 93