Triggering the Oscilloscope 3
5000 Series Oscilloscope User’s Guide 119
1 Press the More key in the Trigger section of the front panel,
rotate the Entry knob until LIN is displayed in the Trigger
softkey.
2 Press the Settings softkey to display LIN trigger menu.
3 Press the Trigger: softkey and choose to trigger on Sync Break
or Frame ID.
Sync (Sync Break) The oscilloscope will trigger on the rising
edge at the Sync Break exit of the LIN single-wire bus signal
that marks the beginning the message frame.
ID (Frame ID) The oscilloscope will trigger when a frame with
an ID equal to the selected value is detected. Use the Entry
knob to select the value for the Frame ID.
Signal
baud rate
Trigger on:
condition
Trigger level
or threshold
Currently selected
signal source
LIN trigger
Return to
previous menu