Making Measurements 4
5000 Series Oscilloscope User’s Guide 191
The following example shows a delay measurement between the
rising edge of channel 1 and the rising edge of channel 2.
Figure 33 Delay Measurement
Phase
Phase is the calculated phase shift from source 1 to source 2,
expressed in degrees. Negative phase shift values indicate that
the rising edge of source 1 occurred after the rising edge of
source 2.
Phase =
Delay
Source 1 Period
-----------------------------------------
X 360
Source 1
Source 2
Delay
Period