86 5000 Series Oscilloscope User’s Guide
2 Front-Panel Controls
Self Test
Pressing Utility&Service&Start Self Test performs a series of
internal procedures to verify that the oscilloscope is operating
properly.
It is recommended you run Self Test:
after experiencing abnormal operation.
for additional information to better describe an oscilloscope
failure.
to verify proper operation after the oscilloscope has been
repaired.
Successfully passing Self Test does not guarantee 100% of the
oscilloscope's functionality. Self Test is designed to provide an
80% confidence level that the oscilloscope is operating properly.