Triggering the Oscilloscope 3
5000 Series Oscilloscope User’s Guide 127
To use Duration triggering
Duration trigger lets you define a pattern, then trigger on a
specified time duration of this logical AND combination of the
channels.
1 Press the More key in the Trigger section of the front panel,
rotate the Entry knob until Duration is displayed in the Trigger
softkey, then press the Settings softkey to display the
Duration trigger menu.
2 For each channel you want to include in the desired pattern,
press the Channel softkey to select the channel.
This is the channel source for the 0, 1, or X condition. As you
press the Channel softkey (or rotate the Entry knob on
mixed-signal oscilloscopes), the channel you select is
highlighted in the Pattern = line directly above the
softkeys and in the upper-right corner of the display next to
Dur”. External trigger may also be specified as a channel in
the pattern when using the 2-channel and 4-channel
oscilloscopes.
Trigger level
or threshold
Selected
channel
Duration
trigger
Channel
select
Logic level
Return to
previous menu
Qualifier
Qualifier time set
Channel 1-4
states
External
Trigger
Vedere la pagina 126