Agilent Technologies Agilent InfiniiVision 5000/6000/7000 Series OscilloscopesUser’s Guide
Contents10 InfiniiVision Oscilloscopes User’s GuideTrigger Holdoff 90Trigger Holdoff Operating Hints 90To set holdoff 90External Trigger Input 922-Ch
4 Triggering100 InfiniiVision Oscilloscopes User’s GuideThe CAN baud rate can be set to:The default baud rate is 1 Mb/s.If the desired baud rate is n
Triggering 4InfiniiVision Oscilloscopes User’s Guide 101Duration TriggerDuration trigger lets you define a pattern, then trigger on a specified time
4 Triggering102 InfiniiVision Oscilloscopes User’s Guide• 1 sets the pattern to 1 (high) on the selected channel. A high is a voltage level that is g
Triggering 4InfiniiVision Oscilloscopes User’s Guide 103• When the out of time range (<>) qualifier is selected, the Entry knob sets the upper
4 Triggering104 InfiniiVision Oscilloscopes User’s GuideEdge TriggerThe Edge trigger type identifies a trigger by looking for a specified edge (slope
Triggering 4InfiniiVision Oscilloscopes User’s Guide 105You can select analog channel 1 or 2, Ext, or Line as the trigger source on any Agilent Infin
4 Triggering106 InfiniiVision Oscilloscopes User’s GuideFlexRay TriggerTo trigger on FlexRay data, you need:• A four-channel MSO6000 or MSO7000 Serie
Triggering 4InfiniiVision Oscilloscopes User’s Guide 107Oscilloscope Controls the VPT1000, Synchronous Mode MonitoringIn addition to decoding FlexRay
4 Triggering108 InfiniiVision Oscilloscopes User’s GuideSetting Up VPT1000/Oscilloscope LAN CommunicationThis is necessary when the oscilloscope cont
Triggering 4InfiniiVision Oscilloscopes User’s Guide 10911 Press the Mode softkey and select Asynchronous.12 Press the Baud and Channel softkeys and
Contents InfiniiVision Oscilloscopes User’s Guide 11I2S Trigger 126LIN Trigger 133Nth Edge Burst Trigger 136Pattern Trigger 138Hex Bus Pattern Trigge
4 Triggering110 InfiniiVision Oscilloscopes User’s Guide5 Press the Mode softkey to select one of these operating modes:• PC — PC controls the VPT100
Triggering 4InfiniiVision Oscilloscopes User’s Guide 111Triggering on FlexRay Frames, Times, Errors, or Events1 To trigger on FlexRay data, you must
4 Triggering112 InfiniiVision Oscilloscopes User’s Guide3 Press the Frame ID softkey, and use the Entry knob to select the frame ID value from All or
Triggering 4InfiniiVision Oscilloscopes User’s Guide 113Triggering on the FlexRay Time ScheduleIn the Time trigger mode, the oscilloscope triggers on
4 Triggering114 InfiniiVision Oscilloscopes User’s Guide4 When Static or Dynamic Segments are selected:a Press the Slot # softkey, and use the Entry
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1155 Press the Errors softkey, and use the Entry knob to select the error type:• All Errors• Cod
4 Triggering116 InfiniiVision Oscilloscopes User’s GuideTriggering on FlexRay Events1 In the Trigger section on the front panel, press the [More] key
Triggering 4InfiniiVision Oscilloscopes User’s Guide 117Glitch or Narrow Pulse CaptureA glitch is a rapid change in the waveform that is usually narr
4 Triggering118 InfiniiVision Oscilloscopes User’s Guide15 ns Narrow Pulse, 20 ms/div, Peak Detect ModeUsing Peak Detect Mode to Find a Glitch1 Conne
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1194 Characterize the glitch with Zoom mode:a Press the [Menu/Zoom] key (or [Main/Delayed] key o
Contents12 InfiniiVision Oscilloscopes User’s GuideStabilizing the Display 174Interpreting the Display 175Graphic Symbols in Softkey Menus 176AutoSca
4 Triggering120 InfiniiVision Oscilloscopes User’s GuideI2C TriggerAn I2C (Inter-IC bus) trigger setup consists of connecting the oscilloscope to the
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1215 Press the Signals softkey to display the I2C Signals Menu.6 Connect an oscilloscope channel
4 Triggering122 InfiniiVision Oscilloscopes User’s Guide11 Return to the previous menu; then, press the Trigger: softkey to select one of the followi
Triggering 4InfiniiVision Oscilloscopes User’s Guide 123• Frame (Start: Addr7: Read: Ack: Data) or Frame (Start: Addr7: Write: Ack: Data) — The oscil
4 Triggering124 InfiniiVision Oscilloscopes User’s GuideFor triggering purposes, a restart is treated as a start condition.12 If you have set the osc
Triggering 4InfiniiVision Oscilloscopes User’s Guide 125If don't care (0xXX) is selected for data, the data will be ignored. The trigger will al
4 Triggering126 InfiniiVision Oscilloscopes User’s GuideI2S Trigger The N5468A (Option SND) I²S trigger and serial decode option adds the ability to
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1276 Press the Signals softkey to display the I2S Signals Menu. A diagram appears showing WS, SC
4 Triggering128 InfiniiVision Oscilloscopes User’s Guide10 Set the trigger levels for the SCLK, WS, and SDATA signals to the middle of the signals:•
Triggering 4InfiniiVision Oscilloscopes User’s Guide 12914 Press the Alignment softkey and rotate the Entry knob to select the desired alignment of t
Contents InfiniiVision Oscilloscopes User’s Guide 13Measurement Definitions 199Time Measurements 199Delay and Phase Measurements 203Voltage Measuremen
4 Triggering130 InfiniiVision Oscilloscopes User’s Guide15 Press the WS Low softkey to select whether WS Low indicates Left or Right channel data. Th
Triggering 4InfiniiVision Oscilloscopes User’s Guide 13120 Press the Trigger softkey and choose a qualifier:• Equal — triggers on the specified audio
4 Triggering132 InfiniiVision Oscilloscopes User’s Guide• Decreasing value — similar to the description above except the trigger occurs on a descreas
Triggering 4InfiniiVision Oscilloscopes User’s Guide 133LIN TriggerLIN (Local Interconnect Network) trigger setup consists of connecting the oscillos
4 Triggering134 InfiniiVision Oscilloscopes User’s Guide5 Press the Signals softkey. The LIN Signals Menu will be displayed.6 Press the Source softke
Triggering 4InfiniiVision Oscilloscopes User’s Guide 13510 Press the Sync Break softkey and select the minimum number of clocks that define a sync br
4 Triggering136 InfiniiVision Oscilloscopes User’s GuideNth Edge Burst TriggerThe Nth Edge Burst trigger lets you trigger on the Nth edge of a burst
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1373 Press the Slope softkey to specify the slope of the edge.4 Press the Idle softkey; then, tu
4 Triggering138 InfiniiVision Oscilloscopes User’s GuidePattern TriggerThe Pattern trigger identifies a trigger condition by looking for a specified
Triggering 4InfiniiVision Oscilloscopes User’s Guide 139• 0 sets the pattern to zero (low) on the selected channel. A low is a voltage level that is
Contents14 InfiniiVision Oscilloscopes User’s GuideSaving Oscilloscope Data 247Selecting a Destination for Your Saved Data248Selecting a File Name 24
4 Triggering140 InfiniiVision Oscilloscopes User’s GuideHex Bus Pattern TriggerYou can specify a bus value on which to trigger. To do this, first def
Triggering 4InfiniiVision Oscilloscopes User’s Guide 141Pulse Width TriggerPulse Width (glitch) triggering sets the oscilloscope to trigger on a posi
4 Triggering142 InfiniiVision Oscilloscopes User’s GuideThe selected pulse polarity is displayed in the upper-right corner of the display. A positive
Triggering 4InfiniiVision Oscilloscopes User’s Guide 143< qualifier time set softkey• When the less than (<) qualifier is selected, the Entry k
4 Triggering144 InfiniiVision Oscilloscopes User’s GuideSequence TriggerSequence trigger lets you trigger the oscilloscope after finding a sequence o
Triggering 4InfiniiVision Oscilloscopes User’s Guide 145As you set stage, term, and channel definitions for the sequence trigger, these settings will
4 Triggering146 InfiniiVision Oscilloscopes User’s GuideDefine the “Find” Stage1 Press the Stage softkey and select Find:.Find: is the first stage in
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1475 If you select an edge term, one channel must be set to a rising edge or a falling edge. All
4 Triggering148 InfiniiVision Oscilloscopes User’s Guide4 If you select a pattern term, each channel in the pattern must be set to a 1 (high), 0 (low
Triggering 4InfiniiVision Oscilloscopes User’s Guide 149• No Reset — Resets on the find condition.• Pattern 1 (or 2) Entered — A pattern is entered o
Contents InfiniiVision Oscilloscopes User’s Guide 15Accessing the Web Interface 279Browser Web Control 280Remote Front Panel 280Remote Programming 282
4 Triggering150 InfiniiVision Oscilloscopes User’s Guide• To set the threshold level for digital channels, press the [D15-D0] key and select Threshol
Triggering 4InfiniiVision Oscilloscopes User’s Guide 151SPI TriggerSerial Peripheral Interface (SPI) trigger setup consists of connecting the oscillo
4 Triggering152 InfiniiVision Oscilloscopes User’s GuideAssign Oscilloscope Channels to SPI SignalsThis explains how to assign oscilloscope channesl
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1536 Press the Clock softkey or turn the Entry knob to select the channel connected to the SPI s
4 Triggering154 InfiniiVision Oscilloscopes User’s Guide8 Press the Data softkey or turn the Entry knob to select the channel that is connected to th
Triggering 4InfiniiVision Oscilloscopes User’s Guide 155• If the framing signal is set to CS (or ~CS), the first clock edge as defined, rising or fal
4 Triggering156 InfiniiVision Oscilloscopes User’s Guideto 32 bits. The data values for the serial string are displayed in the Data string in the wav
Triggering 4InfiniiVision Oscilloscopes User’s Guide 157TV TriggerTV triggering can be used to capture the complicated waveforms of most standard and
4 Triggering158 InfiniiVision Oscilloscopes User’s Guide2 Press the Source softkey and select any analog channel as the TV trigger source.The selecte
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1594 Press the Standard softkey to set the TV standard.The oscilloscope supports triggering on t
Contents16 InfiniiVision Oscilloscopes User’s GuideI2S Serial Decode 310Interpreting I2S Decode 312Interpreting I2S Lister Data 313LIN Serial Decode
4 Triggering160 InfiniiVision Oscilloscopes User’s Guide6 If you select a line # mode, press the Line # softkey, then rotate the Entry knob to select
Triggering 4InfiniiVision Oscilloscopes User’s Guide 161To trigger on a specific line of videoTV triggering requires greater than 1/2 division of syn
4 Triggering162 InfiniiVision Oscilloscopes User’s GuideTo trigger on all sync pulsesTo quickly find maximum video levels, you could trigger on all s
Triggering 4InfiniiVision Oscilloscopes User’s Guide 1633 Press the Mode softkey and select Field1 or Field2.TRIGGERING ON FIELD 1To trigger on all f
4 Triggering164 InfiniiVision Oscilloscopes User’s GuideTRIGGERING ON ALL FIELDSTo trigger on odd or even fieldsTo check the envelope of your video s
Triggering 4InfiniiVision Oscilloscopes User’s Guide 165TRIGGERING ON COLOR FIELD 1 ALTERNATING WITH COLOR FIELD 3If a more detailed analysis is requ
4 Triggering166 InfiniiVision Oscilloscopes User’s GuideHALF-FIELD HOLDOFF TIMEUSING TV HOLDOFF TO SYNCHRONIZE TO COLOR FIELD 1 OR 3 (FIELD 1 MODE)St
Triggering 4InfiniiVision Oscilloscopes User’s Guide 167UART/RS232 TriggerTo trigger on a UART (Universal Asynchronous Receiver/Transmitter) signal c
4 Triggering168 InfiniiVision Oscilloscopes User’s Guide6 Press the Signals softkey to display the UART/RS232 Signals Menu.7 Connect an oscilloscope
Triggering 4InfiniiVision Oscilloscopes User’s Guide 169using the User Baud softkey. You can set the UART baud rate from 1.2 kb/s to 3.0000 Mb/s in i
Contents InfiniiVision Oscilloscopes User’s Guide 1712 Controls and Connectors 349Front Panel 3505000/6000 Series Front Panel (4-channel) 3505000/600
4 Triggering170 InfiniiVision Oscilloscopes User’s Guide15 If you choose a trigger condition that includes “Data” in its description (for example: Rx
Triggering 4InfiniiVision Oscilloscopes User’s Guide 171USB TriggerUSB trigger will trigger on a Start of Packet (SOP), End of Packet (EOP) signal, R
4 Triggering172 InfiniiVision Oscilloscopes User’s GuideAs you press the D+ or D– softkey (or rotate the Entry knob), the D+ and D– label for the sou
173Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 5Displaying Tips for Displaying Waveforms 174Stabilizing the Display 174Inte
5 Displaying174 InfiniiVision Oscilloscopes User’s GuideTips for Displaying WaveformsThe easiest way to display waveforms is to connect the probes to
Displaying 5InfiniiVision Oscilloscopes User’s Guide 175Interpreting the DisplayThe oscilloscope display contains acquired waveforms, setup informati
5 Displaying176 InfiniiVision Oscilloscopes User’s GuideGraphic Symbols in Softkey MenusSoftkeys are the six keys directly below the display. The are
Displaying 5InfiniiVision Oscilloscopes User’s Guide 177AutoScalePress the [AutoScale] key to quickly display the input signals.How AutoScale worksAu
5 Displaying178 InfiniiVision Oscilloscopes User’s GuideSpecifying the Channels Displayed After AutoScaleThe Channels softkey selection determines wh
Displaying 5InfiniiVision Oscilloscopes User’s Guide 179Pan and ZoomThe ability to pan (move horizontally) and zoom (expand or compress horizontally)
Contents18 InfiniiVision Oscilloscopes User’s GuideTo change the displayed size of the digital channels 377To switch a single channel on or off 377To
5 Displaying180 InfiniiVision Oscilloscopes User’s Guide3 Turn the Horizontal Position (Delay Time) knob () to pan horizontally and turn the channel’
Displaying 5InfiniiVision Oscilloscopes User’s Guide 181Amplitude Modulation with Noise Shown at 40% IntensityGrid IntensityTo adjust the grid (grati
5 Displaying182 InfiniiVision Oscilloscopes User’s GuideInfinite PersistenceWith infinite persistence, the oscilloscope updates the display with new
Displaying 5InfiniiVision Oscilloscopes User’s Guide 183Using LabelsYou can define labels and assign them to each analog input channel, or you can tu
5 Displaying184 InfiniiVision Oscilloscopes User’s GuideTo assign a predefined label to a channel1 Press the [Label] key.2 Press the Channel softkey,
Displaying 5InfiniiVision Oscilloscopes User’s Guide 185The channel does not have to be turned on to have a label assigned to it. If the channel is t
5 Displaying186 InfiniiVision Oscilloscopes User’s GuideTo load labels from a text file into the oscilloscope:1 Use a text editor to create each labe
Displaying 5InfiniiVision Oscilloscopes User’s Guide 187This will delete all user-defined labels from the library and set the labels in the library b
5 Displaying188 InfiniiVision Oscilloscopes User’s GuideThe oscilloscope turns vectors on whenever the acquisition system stops. Digital channels on
Displaying 5InfiniiVision Oscilloscopes User’s Guide 189AntialiasingAt slower sweep speeds, the sample rate is reduced and a proprietary display algo
Contents InfiniiVision Oscilloscopes User’s Guide 19Measurement Category 401Measurement Category 401Measurement Category Definitions 401Transient With
5 Displaying190 InfiniiVision Oscilloscopes User’s Guide
191Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 6 Measurements and Math FunctionsList of Automatic Measurements 192Making Au
6 Measurements and Math Functions192 InfiniiVision Oscilloscopes User’s GuideList of Automatic MeasurementsYou can press the [Quick Meas] key and sel
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 193Making Automatic Measurements[Quick Meas] makes automatic measurements o
6 Measurements and Math Functions194 InfiniiVision Oscilloscopes User’s Guide6 To turn off Quick Meas, press the [Quick Meas] key again. The key will
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 195The following statistics are shown: Name of the measurement, current mea
6 Measurements and Math Functions196 InfiniiVision Oscilloscopes User’s GuideTo reset the statistics measurements, press the Reset Statistics softkey
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 1971 From the Quick Meas Menu, press the Settings softkey, then press the T
6 Measurements and Math Functions198 InfiniiVision Oscilloscopes User’s Guide6 Press the Upper softkey, then turn the Entry knob to set the upper mea
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 199Measurement DefinitionsTime MeasurementsThe default lower, middle, and u
Notices© Agilent Technologies, Inc. 2000-2009No part of this manual may be reproduced in any form or by any means (including elec-tronic storage and r
Contents20 InfiniiVision Oscilloscopes User’s GuideTo return the instrument 418Contact us 419Index 421
6 Measurements and Math Functions200 InfiniiVision Oscilloscopes User’s GuideThe measured frequency is normally displayed in 5 digits, but can be dis
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 201ISOLATING AN EVENT FOR FREQUENCY MEASUREMENTPeriodPeriod is the time per
6 Measurements and Math Functions202 InfiniiVision Oscilloscopes User’s Guide+ Width+ Width is the time from the middle threshold of the rising edge
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 203Delay and Phase MeasurementsDelayDelay measures the time difference from
6 Measurements and Math Functions204 InfiniiVision Oscilloscopes User’s GuideThe example below shows a delay measurement between the rising edge of c
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 205The example below shows a phase measurement between the channel 1 and th
6 Measurements and Math Functions206 InfiniiVision Oscilloscopes User’s GuideThe following figure shows the voltage measurement points.AmplitudeThe A
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 207MaximumMaximum is the highest value in the waveform display. The Y curso
6 Measurements and Math Functions208 InfiniiVision Oscilloscopes User’s GuideThe following figure graphically shows the mean and standard deviation.
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 209where: = the standard deviation.N = the number of measurements taken.xi
21Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 1IntroductionModels Covered in this Manual 22Specifications and Characteristi
6 Measurements and Math Functions210 InfiniiVision Oscilloscopes User’s GuideOvershoot and Preshoot MeasurementsPreshootPreshoot is distortion that p
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 211OvershootOvershoot is distortion that follows a major edge transition ex
6 Measurements and Math Functions212 InfiniiVision Oscilloscopes User’s GuidePost Acquisition ProcessingIn addition to changing display parameters af
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 213Cursor MeasurementsCursors are horizontal and vertical markers that indi
6 Measurements and Math Functions214 InfiniiVision Oscilloscopes User’s Guide• Source — Selects a channel, math function, or bus for the cursor measu
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 215In hexadecimal and binary mode, a level can display as 1 (higher than tr
6 Measurements and Math Functions216 InfiniiVision Oscilloscopes User’s GuideThe difference between Y1 and Y2 (Y) is displayed on the dedicated line
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 217Cursor ExamplesCURSORS USED TO MEASURE PULSE WIDTHS OTHER THAN MIDDLE TH
6 Measurements and Math Functions218 InfiniiVision Oscilloscopes User’s GuideExpand the display with Zoom mode, then characterize the event of intere
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 219Press the X1 X2 softkey and move the cursors together to check for pulse
1 Introduction22 InfiniiVision Oscilloscopes User’s GuideAgilent InfiniiVision oscilloscopes feature MegaZoom III technology: • Most responsive deep
6 Measurements and Math Functions220 InfiniiVision Oscilloscopes User’s GuideXY Horizontal ModeThe XY horizontal mode converts the oscilloscope from
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 221SIGNAL CENTERED ON THE DISPLAY4 Press the [Cursors] key.5 Set the Y2 cur
6 Measurements and Math Functions222 InfiniiVision Oscilloscopes User’s GuideCURSORS SET TO CENTER OF SIGNAL7 Calculate the phase difference using th
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 223SIGNALS ARE 90 OUT OF PHASESIGNALS ARE IN PHASE
6 Measurements and Math Functions224 InfiniiVision Oscilloscopes User’s GuideMath FunctionsMath functions can be performed on analog channels. The re
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 2253 Press the Function softkey and select f(t): Displayed. 4 Use the Opera
6 Measurements and Math Functions226 InfiniiVision Oscilloscopes User’s GuideUnitsUnits for each input channel can be set to Volts or Amps using the
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 227MultiplyWhen you select the multiply math function, the voltage values o
6 Measurements and Math Functions228 InfiniiVision Oscilloscopes User’s GuideAdd or SubtractWhen you select add or subtract, the Source 1 and Source
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 229EXAMPLE OF SUBTRACTChannel 1Channel 21 – 2 waveform Subtraction operator
Introduction 1InfiniiVision Oscilloscopes User’s Guide 23Specifications and CharacteristicsPlease see the InfiniiVision oscilloscope data sheets for
6 Measurements and Math Functions230 InfiniiVision Oscilloscopes User’s GuideDifferentiated/dt (differentiate) calculates the discrete time derivativ
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 231A scale unit of U (undefined) will be displayed for scale and offset if
6 Measurements and Math Functions232 InfiniiVision Oscilloscopes User’s GuideIntegrate dt (integrate) calculates the integral of the selected source
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 233INTEGRATE AND SIGNAL OFFSETChannel 1dt waveform Channel 1dt waveform
6 Measurements and Math Functions234 InfiniiVision Oscilloscopes User’s GuideSquare RootSquare root () calculates the square root of the selected so
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 235EXAMPLE OF (SQUARE ROOT).Channel 1 waveform Square root operatorSourc
6 Measurements and Math Functions236 InfiniiVision Oscilloscopes User’s GuideFFT MeasurementFFT is used to compute the fast Fourier transform using a
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 237The following figure illustrates aliasing. This is the spectrum of a 990
6 Measurements and Math Functions238 InfiniiVision Oscilloscopes User’s GuideFFT Operation1 Press the [Math] key, press the Function softkey and sele
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 239• Rectangular — good frequency resolution and amplitude accuracy, but us
1 Introduction24 InfiniiVision Oscilloscopes User’s GuideLicensed Options Many of the following licensed options can be easily installed without retu
6 Measurements and Math Functions240 InfiniiVision Oscilloscopes User’s GuideYou can make peak-to-peak, maximum, minimum, and average dB measurements
Measurements and Math Functions 6InfiniiVision Oscilloscopes User’s Guide 241shape of the window will be the actual limiting factor in the FFTs abili
6 Measurements and Math Functions242 InfiniiVision Oscilloscopes User’s Guide
243Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 7Printing and SavingPrinting the Oscilloscope’s Display 244Quick Print 244To
7 Printing and Saving244 InfiniiVision Oscilloscopes User’s GuidePrinting the Oscilloscope’s DisplayYou can print the complete display, including the
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 245To print the oscilloscope’s display1 Connect a USB printer to either the USB port on
7 Printing and Saving246 InfiniiVision Oscilloscopes User’s Guide• Form Feed — The Form Feed option can be selected to send a form feed command to th
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 247Saving Oscilloscope DataHere are a few points about saving oscilloscope data:• You c
7 Printing and Saving248 InfiniiVision Oscilloscopes User’s GuideSelecting a Destination for Your Saved DataPress [Save/Recall]&Save to access th
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 249If one or more USB mass storage devices are connected to the oscilloscope, you can n
Introduction 1InfiniiVision Oscilloscopes User’s Guide 25See also “Installed Licenses” on page 398.LSS I2C/SPI serial decode option (for 4 channel or
7 Printing and Saving250 InfiniiVision Oscilloscopes User’s Guide4 You can press the Delete Character softkey to delete the current character and shi
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 251• ASCII XY data file— This creates a file of comma-separated variables for each disp
7 Printing and Saving252 InfiniiVision Oscilloscopes User’s GuideThe Invert Graticule Colors option can be used to reduce the amount of black ink it
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 253The length control has no effect on the amount of Serial Decode data saved; all seri
7 Printing and Saving254 InfiniiVision Oscilloscopes User’s GuideFormat (ALB data file format settings in Save Menu)You can acquire data from the osc
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 255RECOMMENDED SIGNAL MAPPINGTo save a waveform and/or setup to a USB mass storage devi
7 Printing and Saving256 InfiniiVision Oscilloscopes User’s GuideTo save a waveform and/or setup to the oscilloscope’s internal memory1 Press the [Sa
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 257File Explorer The File Explorer lets you navigate the file system. Using the File Ex
7 Printing and Saving258 InfiniiVision Oscilloscopes User’s GuideSystem software System software files can be loaded into the oscilloscope. These sof
Printing and Saving 7InfiniiVision Oscilloscopes User’s Guide 259Navigate using Entry knob, select using this key or push Entry knobLoad selected fil
1 Introduction26 InfiniiVision Oscilloscopes User’s GuideNotesBuilt-in Quick HelpA Quick Help system is built into the oscilloscope. Press and hold a
7 Printing and Saving260 InfiniiVision Oscilloscopes User’s Guide
261Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 8Acquisition ModesTo start and stop an acquisition 262To make a single acqui
8 Acquisition Modes262 InfiniiVision Oscilloscopes User’s GuideTo start and stop an acquisition• When you press the [Run/Stop] key, it illuminates in
Acquisition Modes 8InfiniiVision Oscilloscopes User’s Guide 263Memory Depth/Record Length[Run/Stop] versus [Single]When the oscilloscope is running,
8 Acquisition Modes264 InfiniiVision Oscilloscopes User’s GuideTo make a single acquisitionWhen you press the [Single] key it illuminates in yellow a
Acquisition Modes 8InfiniiVision Oscilloscopes User’s Guide 265Acquisition ModesThe InfiniiVision oscilloscopes have the following acquisition modes:
8 Acquisition Modes266 InfiniiVision Oscilloscopes User’s GuidePeak Detect ModeIn Peak Detect mode at slower sweep speeds, minimum and maximum sample
Acquisition Modes 8InfiniiVision Oscilloscopes User’s Guide 267Averaging ModeThe Averaging mode lets you average multiple acquisitions together to re
8 Acquisition Modes268 InfiniiVision Oscilloscopes User’s GuideRANDOM NOISE ON THE DISPLAYED WAVEFORM128 AVERAGES USED TO REDUCE RANDOM NOISESee also
Acquisition Modes 8InfiniiVision Oscilloscopes User’s Guide 269Realtime Sampling OptionRealtime sampling specifies that the oscilloscope produce the
27Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 2Initial SetupPackage Contents 285000 Series Oscilloscope Package Contents 29
8 Acquisition Modes270 InfiniiVision Oscilloscopes User’s GuideTo see the sample rate, press the [Menu/Zoom] key (or [Main/Delayed] key on some oscil
Acquisition Modes 8InfiniiVision Oscilloscopes User’s Guide 271Segmented MemoryYou can purchase the oscilloscope with the segmented memory option fac
8 Acquisition Modes272 InfiniiVision Oscilloscopes User’s GuideIf the signal you are measuring has more than about 1 s of “dead time” (inactivity), c
Acquisition Modes 8InfiniiVision Oscilloscopes User’s Guide 273Measurements, statistics, and infinite persistenceTo perform measurements and view sta
8 Acquisition Modes274 InfiniiVision Oscilloscopes User’s GuideSaving data from segmented memoryYou can save either the currently displayed segment (
275Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 9Web InterfaceSetting up the Oscilloscope’s LAN Connection 276Accessing the
9 Web Interface276 InfiniiVision Oscilloscopes User’s GuideMicrosoft Internet Explorer is the recommended Web browser for communication and control o
Web Interface 9InfiniiVision Oscilloscopes User’s Guide 277To connect to the network using Manual mode1 Get the oscilloscope’s network parameters (ho
9 Web Interface278 InfiniiVision Oscilloscopes User’s GuideHostnameIf you do not know the oscilloscope’s hostname you can reset the hostname as follo
Web Interface 9InfiniiVision Oscilloscopes User’s Guide 279Accessing the Web InterfaceTo access the oscilloscope’s web interface:1 Connect the oscill
2 Initial Setup28 InfiniiVision Oscilloscopes User’s GuidePackage Contents✔ Inspect the shipping container for damage.If your shipping container appe
9 Web Interface280 InfiniiVision Oscilloscopes User’s GuideBrowser Web ControlThe web interface’s Browser Web Control page gives you access to the Re
Web Interface 9InfiniiVision Oscilloscopes User’s Guide 281Scrolling and Monitor ResolutionWhen using a monitor resolution of 1024 x 768 or less on t
9 Web Interface282 InfiniiVision Oscilloscopes User’s GuideRemote ProgrammingTo send remote programming commands to the oscilloscope via the SCPI Com
Web Interface 9InfiniiVision Oscilloscopes User’s Guide 283If you get the “Unable to connect to the instrument. You may need to set the instrument’s
9 Web Interface284 InfiniiVision Oscilloscopes User’s GuideGet ImageTo save (or print) the oscilloscope’s display from the web interface:1 Access the
Web Interface 9InfiniiVision Oscilloscopes User’s Guide 285Identification FunctionThe Identification web interface feature is useful when trying to l
9 Web Interface286 InfiniiVision Oscilloscopes User’s GuideInstrument UtilitiesThe Instrument Utilities page of the web interface lets you:• View ins
Web Interface 9InfiniiVision Oscilloscopes User’s Guide 287Setting a PasswordWhenever you connect the oscilloscope to a LAN, it is good practice to s
9 Web Interface288 InfiniiVision Oscilloscopes User’s Guide4 Enter your desired password, and click Apply Changes.When accessing the password protect
289Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 10Serial Decode/ListerSerial Decode 290Lister 291CAN Serial Decode 293FlexRa
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 295000 Series Oscilloscope Package ContentsInfiniiVision 5000 Series OscilloscopeOscilloscope
10 Serial Decode/Lister290 InfiniiVision Oscilloscopes User’s GuideSerial DecodeAgilent’s hardware accelerated serial decode options can be installed
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 291ListerLister is a powerful tool for investigating protocol failures. You can use L
10 Serial Decode/Lister292 InfiniiVision Oscilloscopes User’s Guide4 Press the [Single] key (in the Run Control group on the front panel) to stop the
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 293CAN Serial Decode1 Connect the CAN signal to one of the oscilloscope’s input chann
10 Serial Decode/Lister294 InfiniiVision Oscilloscopes User’s Guide4 Specify the CAN signal:a Press the Settings softkey to access the CAN Signals Me
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 295The value of the trigger level or digital threshold is displayed in the upper-righ
10 Serial Decode/Lister296 InfiniiVision Oscilloscopes User’s Guide10 If the decode line does not appear on the display, press the Decode softkey to
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 297• Data length code (DLC) appears in blue for data frames and green for remote fram
10 Serial Decode/Lister298 InfiniiVision Oscilloscopes User’s GuideCounters• The FRAMES counter gives the total number of completed remote, data, ove
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 299Interpreting CAN Lister DataIn addition to the standard Time column, the CAN Liste
InfiniiVision Oscilloscopes User’s Guide 3Book MapInitial Setup Connecting to the Device Under TestTriggeringDisplaying Measurements and Math Functio
2 Initial Setup30 InfiniiVision Oscilloscopes User’s Guide6000A Series Oscilloscope Package ContentsPower cord(Based on country of origin)6000A Serie
10 Serial Decode/Lister300 InfiniiVision Oscilloscopes User’s GuideFlexRay Serial DecodeTo trigger on FlexRay data, you need:• A four-channel MSO6000
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 301c In the Serial Decode Menu, press the Mode softkey.d Use the Entry knob (or repea
10 Serial Decode/Lister302 InfiniiVision Oscilloscopes User’s GuideInterpreting FlexRay Frame DecodeThe FlexRay decode display shows frame decoding (
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 303Interpreting FlexRay Time DecodeGlobal Time-Schedule Decoding(top row of decode di
10 Serial Decode/Lister304 InfiniiVision Oscilloscopes User’s GuideThe totalizer is running, counting frames and calculating percentages, even when t
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 305Interpreting FlexRay Lister DataIn addition to the standard Time column, the FlexR
10 Serial Decode/Lister306 InfiniiVision Oscilloscopes User’s GuideI2C Serial Decode1 Select the I2C serial decode mode:a Press the [Acquire] key.b P
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 307a Press the Settings softkey to access the I2C Signals Menu.b Press the SCL softke
10 Serial Decode/Lister308 InfiniiVision Oscilloscopes User’s GuideInterpreting I2C Decode• Angled waveforms show an active bus (inside a packet/fram
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 309• Aliased bus values (undersampled or indeterminate) are drawn in red.• Unknown bu
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 316000A Option BAT Oscilloscope Package ContentsAC/DC power adapterFront-panel cover6000A Ser
10 Serial Decode/Lister310 InfiniiVision Oscilloscopes User’s GuideI2S Serial Decode1 Select the I2S serial decode mode:a Press the [Acquire] key.b P
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 3112 Press the Settings softkey to open the I2S Settings Menu.a Press the Signals sof
10 Serial Decode/Lister312 InfiniiVision Oscilloscopes User’s GuideInterpreting I2S Decode• Angled waveforms show an active bus (inside a packet/fram
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 313Interpreting I2S Lister DataIn addition to the standard Time column, the I2S Liste
10 Serial Decode/Lister314 InfiniiVision Oscilloscopes User’s GuideLIN Serial Decode1 Select the LIN serial decode mode:a Press the [Acquire] key.b P
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 3153 Specify the LIN signal:a Press the Settings softkey to access the LIN Signals Me
10 Serial Decode/Lister316 InfiniiVision Oscilloscopes User’s GuideIf the desired baud rate is not shown in the list, select User Defined. You can se
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 3175 If the oscilloscope is stopped, press the [Run/Stop] key to acquire and decode d
10 Serial Decode/Lister318 InfiniiVision Oscilloscopes User’s Guide• For LIN 1.3, the checksum appears in blue if correct, or red if incorrect. The c
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 319Interpreting LIN Lister DataIn addition to the standard Time column, the LIN Liste
2 Initial Setup32 InfiniiVision Oscilloscopes User’s Guide6000L Series Oscilloscope Package Contents6000L Series OscilloscopeOscilloscope Probes10073
10 Serial Decode/Lister320 InfiniiVision Oscilloscopes User’s GuideSPI Serial Decode1 Select the SPI serial decode mode:a Press the [Acquire] key.b P
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 3212 Specify the SPI signals:a Press the Settings softkey to access the SPI Signals M
10 Serial Decode/Lister322 InfiniiVision Oscilloscopes User’s Guideb Press the Clock softkey, and use the Entry knob to select the channel probing th
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 3234 If the decode line does not appear on the display, press the Decode softkey to t
10 Serial Decode/Lister324 InfiniiVision Oscilloscopes User’s Guide• Decoded text is truncated at the end of the associated frame when there is insuf
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 325UART/RS232 Serial Decode1 Select the UART/RS232 serial decode mode:a Press the [Ac
10 Serial Decode/Lister326 InfiniiVision Oscilloscopes User’s Guide2 Specify the UART/RS232 signals:a Press the Settings softkey to access the UART/R
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 327b Press the Signals softkey, and select the oscilloscope channels that you have co
10 Serial Decode/Lister328 InfiniiVision Oscilloscopes User’s Guide8 Make sure the trigger or threshold levels are set to the middle of the signals:•
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 329Interpreting UART/RS232 Decode• Angled waveforms show an active bus (inside a pack
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 337000A Series Oscilloscope Package ContentsPower cord(Based on country of origin)*Digital Pr
10 Serial Decode/Lister330 InfiniiVision Oscilloscopes User’s Guide• A parity error will cause the associated data word to be shown in red, which inc
Serial Decode/Lister 10InfiniiVision Oscilloscopes User’s Guide 331Interpreting UART/RS232 Lister DataIn addition to the standard Time column, the UA
10 Serial Decode/Lister332 InfiniiVision Oscilloscopes User’s Guide
333Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 11Mask TestTo create a mask from a “golden” waveform (Automask) 334Setup Opt
11 Mask Test334 InfiniiVision Oscilloscopes User’s GuideTo create a mask from a “golden” waveform (Automask)A golden waveform meets all chosen parame
Mask Test 11InfiniiVision Oscilloscopes User’s Guide 3357 Press the Create Mask softkey. The mask is created and testing begins. Whenever the Create
11 Mask Test336 InfiniiVision Oscilloscopes User’s GuideSetup OptionsRun UntilThe Run Until softkey lets you specify a condition on which to terminat
Mask Test 11InfiniiVision Oscilloscopes User’s Guide 337Save The oscilloscope will save the screen image based on the settings you have chosen. See “
11 Mask Test338 InfiniiVision Oscilloscopes User’s GuideMask Test Trigger OutputThe oscilloscope’s trigger output (located on the rear panel) can be
Mask Test 11InfiniiVision Oscilloscopes User’s Guide 339Mask StatisticsFrom the Mask Test Menu, press the Statistics softkey to enter the Mask Statis
2 Initial Setup34 InfiniiVision Oscilloscopes User’s GuideAccessories Available Model DescriptionN2918A 6000/7000 Series Oscilloscope Evaluation KitN
11 Mask Test340 InfiniiVision Oscilloscopes User’s GuideReset StatisticsPress this softkey to reset the statistics counters to zero. Note that statis
Mask Test 11InfiniiVision Oscilloscopes User’s Guide 341To manually modify a mask fileYou can manually modify a mask file that you created using the
11 Mask Test342 InfiniiVision Oscilloscopes User’s GuideMask Violation RegionsUp to 8 regions can be defined for a mask. They can be numbered 1-8. Th
Mask Test 11InfiniiVision Oscilloscopes User’s Guide 343• X• Y1• Y2These four parameters are defined in the Oscilloscope Setup portion of the mask f
11 Mask Test344 InfiniiVision Oscilloscopes User’s GuideBuilding a Mask FileThe following mask uses all eight mask regions. The most difficult part o
Mask Test 11InfiniiVision Oscilloscopes User’s Guide 345 -15.00, 0.500 -15.00, 1.500 -12.50, 1.500/* Region Number */ 3/* Number of ve
11 Mask Test346 InfiniiVision Oscilloscopes User’s Guide/* Region Number */ 8/* Number of vertices */ 4 -12.50, MIN -10.00, -1.750 10
Mask Test 11InfiniiVision Oscilloscopes User’s Guide 347The mask file allows greater resolution than the 256 X 1000 database. Some quantization of da
11 Mask Test348 InfiniiVision Oscilloscopes User’s Guide
349Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 12Controls and ConnectorsFront Panel 3505000/6000 Series Front Panel (4-chan
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 35W2637A LPDDR BGA probe, x16, 404 MHz, 100 ohm input impedanceW2638A LPDDR BGA probe, x32, 4
12 Controls and Connectors350 InfiniiVision Oscilloscopes User’s GuideFront Panel5000/6000 Series Front Panel (4-channel) Click the text for the item
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 3515000/6000 Series Front Panel (2-Channel, differences only)The differences betwe
12 Controls and Connectors352 InfiniiVision Oscilloscopes User’s Guide6000L Series Front and Rear PanelClick the text for the item’s description and
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 3537000 Series Front Panel (4-Channel) Click the text for the item’s description a
12 Controls and Connectors354 InfiniiVision Oscilloscopes User’s Guide7000 Series Front Panel (2-Channel, differences only)The differences between th
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 355Front Panel Control and Connector Descriptions1. Power Switch Press once to swi
12 Controls and Connectors356 InfiniiVision Oscilloscopes User’s Guide6. [Math] Key The [Math] key provides access to add, subtract, multiply, differ
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 357If a trace is repositioned over an existing trace the indicator at the left edg
12 Controls and Connectors358 InfiniiVision Oscilloscopes User’s GuideHorizontal Sweep Speed Control Turn the knob in the Horizontal section that is
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 35923. Softkeys The functions of these keys change based upon the menus shown on t
2 Initial Setup36 InfiniiVision Oscilloscopes User’s GuideYou can find these items at www.agilent.com or at www.parts.agilent.com.For information on
12 Controls and Connectors360 InfiniiVision Oscilloscopes User’s GuideRear Panel5000 Series Rear Panel (4-Channel)Click the text for the item’s descr
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 3615000 Series Rear Panel (2-Channel)The 2-channel model rear panel is the same as
12 Controls and Connectors362 InfiniiVision Oscilloscopes User’s Guide6000 Series Rear Panel (4-Channel) Click the text for the item’s description an
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 3636000 Series Rear Panel (2-Channel)The 2-channel model rear panel is the same as
12 Controls and Connectors364 InfiniiVision Oscilloscopes User’s Guide6000 Series Option BAT Rear Panel (4-Channel)Click the text for the item’s desc
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 3657000 Series Rear Panel (4-Channel) Click the text for the item’s description an
12 Controls and Connectors366 InfiniiVision Oscilloscopes User’s Guide7000 Series Rear Panel (2-Channel)The 2-channel model rear panel is the same as
Controls and Connectors 12InfiniiVision Oscilloscopes User’s Guide 367Rear Panel Control and Connector Descriptions1. External Trigger Input See “Ext
12 Controls and Connectors368 InfiniiVision Oscilloscopes User’s Guide13. Ground Terminal Option BAT oscilloscopes: Connect the oscilloscope to earth
369Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 13Digital ChannelsTo connect the digital probes to the device under test 370
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 37The tilt tabs (underneath the oscilloscope) can be positioned as shown in the center pictur
13 Digital Channels370 InfiniiVision Oscilloscopes User’s GuideTo connect the digital probes to the device under testThe digital channels are enabled
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 3714 Connect the ground lead on each set of channels (each pod), using a probe grabber. T
13 Digital Channels372 InfiniiVision Oscilloscopes User’s Guide6 Connect the grabber to a node in the circuit you want to test.7 For high-speed signa
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 3738 Repeat steps 3 through 6 until you have connected all points of interest. Acquiring
13 Digital Channels374 InfiniiVision Oscilloscopes User’s GuideTo display digital channels using AutoScaleWhen signals are connected to the digital c
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 375the instrument to the factory default configuration by pressing the [Save/Recall] key,
13 Digital Channels376 InfiniiVision Oscilloscopes User’s GuideInterpreting the digital waveform displayThe following figure shows a typical display
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 377To change the displayed size of the digital channels1 Press the [D15-D0] key.2 Press t
13 Digital Channels378 InfiniiVision Oscilloscopes User’s GuideTo change the logic threshold for digital channels1 Press the [D15-D0] key so that the
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 379If a trace is repositioned over an existing trace the indicator at the left edge of th
2 Initial Setup38 InfiniiVision Oscilloscopes User’s Guide2 Without releasing the hubs, rotate the handle to the desired position. Then release the h
13 Digital Channels380 InfiniiVision Oscilloscopes User’s GuideThe buses are shown at the bottom of the display.Using cursors to read bus valuesTo re
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 3814 Use the Entry knob and the X1 and X2 softkeys to position the cursors where you want
13 Digital Channels382 InfiniiVision Oscilloscopes User’s GuideBus values are displayed when using Pattern triggerThe bus values are also displayed w
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 383Digital channel signal fidelity: Probe impedance and groundingWhen using the mixed-sig
13 Digital Channels384 InfiniiVision Oscilloscopes User’s GuideHIGH-FREQUENCY PROBE EQUIVALENT CIRCUIT The impedance plots for the two models are sho
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 385Probe GroundingA probe ground is the low-impedance path for current to return to the s
13 Digital Channels386 InfiniiVision Oscilloscopes User’s Guidereconstructed waveforms, they do not show ringing and perturbations. You will not find
Digital Channels 13InfiniiVision Oscilloscopes User’s Guide 387To replace digital probe leadsIf you need to remove a probe lead from the cable, inser
13 Digital Channels388 InfiniiVision Oscilloscopes User’s Guide
389Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 14UtilitiesTo restore the oscilloscope to its default configuration 390To pe
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 393 Rock the oscilloscope back so that it rests securely on its feet.To retract the feet1 Til
14 Utilities390 InfiniiVision Oscilloscopes User’s GuideTo restore the oscilloscope to its default configuration• Press the [Save/Recall] key, then p
Utilities 14InfiniiVision Oscilloscopes User’s Guide 391To perform service functions• Press the [Utility]&Service to display the Service Menu.The
14 Utilities392 InfiniiVision Oscilloscopes User’s GuideUser CalibrationPerform user-calibration:• Each year or after 2000 hours of operation.• If th
Utilities 14InfiniiVision Oscilloscopes User’s Guide 393For a 2-channel oscilloscope, connect a BNC tee to the equal length cables. Then connect a BN
14 Utilities394 InfiniiVision Oscilloscopes User’s Guide3 Connect a BNC cable (40 inches maximum) from the TRIG OUT connector on the rear panel to th
Utilities 14InfiniiVision Oscilloscopes User’s Guide 395Self TestPressing [Utility]&Service&Start Self Test performs a series of internal pro
14 Utilities396 InfiniiVision Oscilloscopes User’s Guide
397Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 15ReferenceAbout Oscilloscope 398Software and Firmware Updates 398Installed
15 Reference398 InfiniiVision Oscilloscopes User’s GuideAbout OscilloscopePress [Utility]&Service&About Oscilloscope to display information a
Reference 15InfiniiVision Oscilloscopes User’s Guide 399•LSS — Low Speed Serial Decode. This license provides serial decode of I2C and SPI buses.•mem
4 InfiniiVision Oscilloscopes User’s GuideIn This User’s Guide…This guide shows you how to use the InfiniiVision 5000/6000/7000 Series oscilloscopes.
2 Initial Setup40 InfiniiVision Oscilloscopes User’s GuideTo install the optional front panel overlay5000 Series OverlayWhen Option AB0 (Traditional
15 Reference400 InfiniiVision Oscilloscopes User’s GuideSecure Environment Mode OptionSecure environment mode is compliant to National Industrial Sec
Reference 15InfiniiVision Oscilloscopes User’s Guide 401Measurement CategoryMeasurement CategoryThe InfiniiVision oscilloscopes are intended to be us
15 Reference402 InfiniiVision Oscilloscopes User’s GuideTransient Withstand CapabilityCAUTIONMaximum input voltage at analog inputs and 2-channel ext
Reference 15InfiniiVision Oscilloscopes User’s Guide 403Environmental ConditionsEnvironment Indoor use only.Ambient temperature Operating 0 °C to +5
15 Reference404 InfiniiVision Oscilloscopes User’s GuideUsing the 10 MHz reference clockThe 10 MHz REF BNC connector on the rear panel is provided so
Reference 15InfiniiVision Oscilloscopes User’s Guide 405A black locked padlock icon will appear at the top of the display.If the externally supplied
15 Reference406 InfiniiVision Oscilloscopes User’s GuideTo synchronize the timebase of two or more instrumentsThe oscilloscope can output it’s 10 MHz
Reference 15InfiniiVision Oscilloscopes User’s Guide 407Binary Data (.bin) FormatThe binary data format stores waveform data in binary format and pro
15 Reference408 InfiniiVision Oscilloscopes User’s GuideAgilent provides the .m files, which need to be copied into the work directory for MATLAB. Th
Reference 15InfiniiVision Oscilloscopes User’s Guide 409Number of Waveform Buffers A 32-bit integer that is the number of waveform buffers required t
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 416000 Series OverlayLarge label: Remove the protective backing to expose the adhesive, align
15 Reference410 InfiniiVision Oscilloscopes User’s GuideY Units A 32-bit integer that identifies the unit of measure for Y values in the acquired dat
Reference 15InfiniiVision Oscilloscopes User’s Guide 411Buffer Size A 32-bit integer that is the size of the buffer required to hold the data points.
15 Reference412 InfiniiVision Oscilloscopes User’s GuideExamples of Binary FilesSingle Acquisition Multiple Analog ChannelsThe following picture show
Reference 15InfiniiVision Oscilloscopes User’s Guide 413Single Acquisition All Pods Logic ChannelsThe following picture shows a binary file of a sing
15 Reference414 InfiniiVision Oscilloscopes User’s GuideFile Header12 bytesWaveform Header 1140 bytesWaveform DataHeader 112 bytesVoltage Data 1buffe
Reference 15InfiniiVision Oscilloscopes User’s Guide 415CSV and ASCII XY filesCSV and ASCII XY file structureIn CSV or ASCII XY format the Length con
15 Reference416 InfiniiVision Oscilloscopes User’s GuideVPT1000 IP address and firmware updatesChanging the VPT1000 network settingsTo change the Net
Reference 15InfiniiVision Oscilloscopes User’s Guide 417AcknowledgementsRealVNC is licensed under the GNU General Public License. Copyright (C) 2002-
15 Reference418 InfiniiVision Oscilloscopes User’s GuideTo return the instrumentBefore shipping the oscilloscope to Agilent Technologies, contact you
Reference 15InfiniiVision Oscilloscopes User’s Guide 419Contact usAmericasCanada (877) 894-4414Latin America 305 269 7500United States (800) 829-4444
2 Initial Setup42 InfiniiVision Oscilloscopes User’s Guide7000 Series OverlayA front panel overlay which labels the controls is available in Simplifi
15 Reference420 InfiniiVision Oscilloscopes User’s Guide
InfiniiVision Oscilloscopes User’s Guide 421IndexSymbols(-)Width measurement, 202(+)Width measurement, 202Numerics1 M ohm input impedance, 6110 MHz r
Index 422 InfiniiVision Oscilloscopes User’s Guidecalibration protect switch5000 Series, 3606000 Series, 3626000-BAT Series, 362, 3647000 Series, 365c
Index InfiniiVision Oscilloscopes User’s Guide 423edge triggering, 104EEPROM data read, I2C trigger, 122energy of a pulse, 232Entry knob, 53, 176, 358
Index 424 InfiniiVision Oscilloscopes User’s GuideJJapanese GUI, 56Kkeys, 535000/6000 Series, 350, 3516000L Series, 3527000 Series, 353, 354Knob softk
Index InfiniiVision Oscilloscopes User’s Guide 425normal acquire mode, 265normal horizontal mode, 64normal mode, 265Normal trigger mode, 84, 85, 357Nt
Index 426 InfiniiVision Oscilloscopes User’s Guiderunt pulses, 201Ssafety warning, 76sample clock, 10 MHz, 404sample rate, 22, 23, 266, 269current rat
Index InfiniiVision Oscilloscopes User’s Guide 427triggercoupling, 87definition, 83external, 92general information, 83holdoff, 90list of trigger types
Index 428 InfiniiVision Oscilloscopes User’s GuideZZ-axis blanking, 69, 222zoom and pan, 179Zoom mode, 66
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 43To mount the oscilloscope in a rackThe InfiniiVision oscilloscopes can be mounted in an Ele
2 Initial Setup44 InfiniiVision Oscilloscopes User’s Guide1 Loosely attach the Front Extender Supports to the Rear Extender Supports with four (4) cl
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 45a Attach a Rack Mount Extender to the left side of the oscilloscope using two (2) of the M3
2 Initial Setup46 InfiniiVision Oscilloscopes User’s Guide6000L Mounting Bracket DimensionsThe following information is provided in case you want to
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 47Ventilation RequirementsThe air intake and exhaust areas must be free from obstructions. Un
2 Initial Setup48 InfiniiVision Oscilloscopes User’s GuidePower Requirements5000, 6000A (without Option BAT), and 7000 Line voltage, frequency, and p
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 49Power-on the 5000/6000/7000 Series Oscilloscope1 Connect the power cord to the rear of the
InfiniiVision Oscilloscopes User’s Guide 59Web InterfaceSetting up the I/O port, establishing LAN connection, using the oscilloscope’s web interface.
2 Initial Setup50 InfiniiVision Oscilloscopes User’s GuideThe battery will last approximately 1.75 hours before requiring recharging. Operating time
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 513 Press the power switch.The power switch is located on the lower left corner of the front
2 Initial Setup52 InfiniiVision Oscilloscopes User’s GuideYou can use the oscilloscope while the battery is being charged.Operating with the Automoti
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 53Keys, Softkeys, and the Entry KnobOn the front panel, “key” refers to any key (button) you
2 Initial Setup54 InfiniiVision Oscilloscopes User’s GuideVerifying Basic Oscilloscope OperationIf you have a 6000L Series oscilloscope, you need to
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 55Quick HelpTo view Quick Help1 Press and hold the key or softkey for which you would like to
2 Initial Setup56 InfiniiVision Oscilloscopes User’s GuideQuick Help Languages To choose a Quick Help language in the oscilloscope:1 Press [Utility],
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 57To set the clockThe Clock Menu lets you set the current date and time of day (24-hour forma
2 Initial Setup58 InfiniiVision Oscilloscopes User’s GuideTo set up the screen saverThe oscilloscope can be configured to turn on a display screen sa
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 595 To view the normal display after the screen saver has started, press any key or turn any
6 InfiniiVision Oscilloscopes User’s Guide
2 Initial Setup60 InfiniiVision Oscilloscopes User’s GuidePressing an analog channel key displays the channel’s menu and turns the display of the cha
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 61Channel Coupling and Input Impedance Selection3 Press the channel’s on/off key, then press
2 Initial Setup62 InfiniiVision Oscilloscopes User’s Guide• 1M Ohm mode is for use with many passive probes and for general-purpose measurements. The
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 63• Probe Attenuation Factor— This is set automatically if the oscilloscope can identify the
2 Initial Setup64 InfiniiVision Oscilloscopes User’s GuideTo set up the Horizontal time base 1 Press the [Menu/Zoom] key (or [Main/Delayed] key on so
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 65The Vernier softkey lets you change the sweep speed in smaller increments with the time/div
2 Initial Setup66 InfiniiVision Oscilloscopes User’s GuideAll events displayed left of the trigger point happened before the trigger occurred, and th
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 67When the horizontal mode is set to Normal, the horizontal position control (delay control)
2 Initial Setup68 InfiniiVision Oscilloscopes User’s GuideRoll mode• Press the [Menu/Zoom] key (or [Main/Delayed] key on some oscilloscopes), then pr
Initial Setup 2InfiniiVision Oscilloscopes User’s Guide 69Knob softkeyThe “Knob “softkey is used in conjunction with Segmented Memory. You can set t
InfiniiVision Oscilloscopes User’s Guide 7ContentsBook Map 3In This User’s Guide… 41 Introduction 21Models Covered in this Manual 22Specifications an
2 Initial Setup70 InfiniiVision Oscilloscopes User’s Guide
71Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 3Connecting to the Device Under TestAnalog Input Impedance (50 Ohm or 1 MOhm)
3 Connecting to the Device Under Test72 InfiniiVision Oscilloscopes User’s GuideAnalog Input Impedance (50 Ohm or 1 MOhm)The input impedance of most
Connecting to the Device Under Test 3InfiniiVision Oscilloscopes User’s Guide 73Passive Probes All InfiniiVision oscilloscopes recognize passive prob
3 Connecting to the Device Under Test74 InfiniiVision Oscilloscopes User’s GuideActive Probes Most Agilent active probes are compatible with the Auto
Connecting to the Device Under Test 3InfiniiVision Oscilloscopes User’s Guide 75Active Probes for 6000 Series 100 MHz Bandwidth ModelsThe following a
3 Connecting to the Device Under Test76 InfiniiVision Oscilloscopes User’s GuideCAUTIONMaximum input voltage at analog inputs: CAT I 300 Vrms, 400 Vp
Connecting to the Device Under Test 3InfiniiVision Oscilloscopes User’s Guide 77Compensating Passive ProbesEach oscilloscope passive probe must be co
3 Connecting to the Device Under Test78 InfiniiVision Oscilloscopes User’s GuideCalibrating ProbesSome probes, such as the 10073C, 10074C, and 1165A
Connecting to the Device Under Test 3InfiniiVision Oscilloscopes User’s Guide 79• 10:1 (+20 dB Atten) single-ended browser• 10:1 (+20 dB Atten) diffe
Contents8 InfiniiVision Oscilloscopes User’s GuideTo mount the oscilloscope in a rack 43To mount the 5000 or 6000 Series oscilloscope in a rack 43To
3 Connecting to the Device Under Test80 InfiniiVision Oscilloscopes User’s GuideManually Setting the Probe Attenuation FactorIf you connect a probe t
81Agilent InfiniiVision 5000/6000/7000 Series OscilloscopeUser’s Guide 4TriggeringTriggering - General Information 83Trigger Mode and Coupling Menu
4 Triggering82 InfiniiVision Oscilloscopes User’s GuideThis chapter provides instructions for setting up triggers for various signal types. Setups ca
Triggering 4InfiniiVision Oscilloscopes User’s Guide 83Triggering - General InformationA triggered waveform is one in which the oscilloscope begins t
4 Triggering84 InfiniiVision Oscilloscopes User’s GuideTrigger Mode and Coupling Menu• Press the [Mode/Coupling] key in the Trigger section of the fr
Triggering 4InfiniiVision Oscilloscopes User’s Guide 85Auto ModeUse the auto trigger modes for signals other than low-repetitive-rate signals and for
4 Triggering86 InfiniiVision Oscilloscopes User’s Guidepre-trigger buffer, the trigger will not be found. If you use Normal mode and wait for the tri
Triggering 4InfiniiVision Oscilloscopes User’s Guide 87Trigger Coupling1 Press the [Mode/Coupling] key.2 Press the Coupling softkey, then select DC,
4 Triggering88 InfiniiVision Oscilloscopes User’s GuideHF RejectHF Reject adds a 50 kHz low-pass filter in the trigger path to remove high frequency
Triggering 4InfiniiVision Oscilloscopes User’s Guide 89LF RejectLow-frequency reject (LF reject) adds a high-pass filter with the 3-dB point at 50 kH
Contents InfiniiVision Oscilloscopes User’s Guide 9AutoProbe Interface 72Passive Probes 73Active Probes 74Active Probes for 6000 Series 100 MHz Bandwi
4 Triggering90 InfiniiVision Oscilloscopes User’s GuideTrigger HoldoffTrigger Holdoff Operating HintsHoldoff keeps a trigger from occurring until aft
Triggering 4InfiniiVision Oscilloscopes User’s Guide 91By setting the Holdoff, you can synchronize triggers. The oscilloscope will trigger on one edg
4 Triggering92 InfiniiVision Oscilloscopes User’s GuideExternal Trigger InputThe External Trigger input can be used as a source in several of the tri
Triggering 4InfiniiVision Oscilloscopes User’s Guide 93Range The input voltage range can be set to 1.0 Volts or 8.0 Volts. When in current mode, the
4 Triggering94 InfiniiVision Oscilloscopes User’s Guide4-Channel Oscilloscope External Trigger InputInput Impedance The external trigger input impeda
Triggering 4InfiniiVision Oscilloscopes User’s Guide 95Trigger OutputYou can select one of the following signals to be output at the TRIG OUT connect
4 Triggering96 InfiniiVision Oscilloscopes User’s GuideTrigger TypesThe oscilloscope lets you synchronize the display to the actions of the device un
Triggering 4InfiniiVision Oscilloscopes User’s Guide 97CAN TriggerThe Controller Area Network (CAN) trigger allows triggering on CAN version 2.0A and
4 Triggering98 InfiniiVision Oscilloscopes User’s Guide5 Press the Settings softkey to display CAN Trigger Menu.6 Repeatedly press and release the Tr
Triggering 4InfiniiVision Oscilloscopes User’s Guide 997 Press the Signals softkey to enter the CAN Signals Menu.8 Press the Source softkey to select