Agilent Technologies InfiniiVision 5000 Series Guida Utente Pagina 344

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 343
11 Mask Test
344 InfiniiVision Oscilloscopes User’s Guide
Building a Mask File
The following mask uses all eight mask regions. The most difficult part of creating a mask
file is normalizing the X and Y values from the time and voltage values. This example
shows an easy way to convert voltage and time to normalized X and Y values in the mask
file.
The following mask file produced the mask shown above:
MASK_FILE_548XX
"All Regions"
/* Region Number */ 1
/* Number of vertices */ 4
-12.50, MAX
-10.00, 1.750
10.00, 1.750
12.50, MAX
/* Region Number */ 2
/* Number of vertices */ 5
-10.00, 1.000
-12.50, 0.500
Vedere la pagina 343
1 2 ... 339 340 341 342 343 344 345 346 347 348 349 ... 427 428

Commenti su questo manuale

Nessun commento