4 Triggering
106 InfiniiVision Oscilloscopes User’s Guide
FlexRay Trigger
To trigger on FlexRay data, you need:
A four-channel MSO6000 or MSO7000 Series mixed signal oscilloscope.
Option FRS (N5432B) FlexRay triggering and decode, which includes:
Agilent VPT1000 protocol analyzer module.
MSO communications cable.
Option FRS lets you:
Control the VPT1000 (without a PC).
Trigger on FlexRay bus frames, synchronous timing slots/segments, or error
conditions.
Display decoded FlexRay bus data and synchronous timing.
Modes of VPT1000 Control/Operation
The oscilloscope and VPT1000 can be used in one of three ways:
Oscilloscope controls the VPT1000, asynchronous mode monitoring (LAN connection
required).
Oscilloscope controls the VPT1000, synchronous mode monitoring (LAN connection
required).
PC controls the VPT1000. (Requires appropriate FlexRay protocol analysis software.)
When the oscilloscope controls the VPT1000, the system is a standalone, time-correlated
measurement system (no external PC required).
Oscilloscope Controls the VPT1000, Asynchronous Mode Monitoring
Asynchronous analysis provides decoding of FlexRay frames based on user-selected
channel and baud rate. These can be set in the oscilloscopes VPT1000 Menu.
NOTE
To display FlexRay serial decode, see page 300.