4 Triggering
116 InfiniiVision Oscilloscopes User’s Guide
Triggering on FlexRay Events
1 In the Trigger section on the front panel, press the [More] key.
2 Press the Trigger softkey, and use the Entry knob to select FlexRay.
3 Press the Settings softkey to access the FlexRay Trigger Menu.
4 In the FlexRay Trigger Menu, press the Trigg er softkey, and use the Entry knob to
select Event.
5 Press the Source softkey, and use the Entry knob to select the event source input
channel.
6 Press the Event softkey, and use the Entry knob to select the event type:
Wake-up
TSS - Transmission Start Sequence
FSS - Frame Start Sequence
FES/DTS - Frame End or Dynamic Trailing Sequence
7 Press the Autoset for Event softkey.
This automatically configures oscilloscope settings (as shown on the display) for the
selected event trigger.
Trigger
condition
VPT1000
Menu
Return to
previous menu
Event source
Event type
Auto setup for
event type
Vedere la pagina 115