Agilent Technologies Ininiium 9000 Specifiche Pagina 109

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 108
Chapter 5: Troubleshooting
To check the LEDs
109
To check the LEDs
If any of the LEDs are not working on the front panel, the first step is to trouble shoot the acquisition
assembly to make sure it is working. If it is working then replace the keyboard assembly and see if the
LEDs will illuminate. If the LEDs will not illuminate, the last step is to reload the oscilloscope software.
When reassembling the oscilloscope all boards that were replaced, but turned out not to be causing the
problem should be reinstalled in the oscilloscope.