Chapter 6: Replacing Assemblies
To set the calibration factors after replacing the acquisition board
141
adapter. The other end of the BNC cable will end up being connected to various ports during the
calibration. The calibration instructions on the oscilloscope will instruct you in this regard.
c Pull down the Utilities menu and select Calibration.
d Uncheck to Cal Memory Protect box to allow calibration.
e Click on Start to start the calibration procedure.
Follow the on-screen instructions as the calibration proceeds.