Chapter 4: Testing Performance
Time Scale Accuracy (TSA)
74
15
Compare the measured time scale error to the TSA specification. Note that the measurement
result should exceed the specified accuracy by at least 0.1 ppm in order to guarantee compliance
to the specification.
16 Record the results in the Performance Test Record.