Eye-Diagram Analyzer Reference
Agilent 70820A User-Corrections
Agilent 70820A User-Corrections
User corrections provide a convenient method for modifying the display of
input signals. Some of the applications for user-corrections include the follow-
ing:
Filtering input signals to simulate the effects of circuit design.
Compensating for frequency and phase response of external probes.
Compensating for frequency and phase response of external test setups.
Compensating for the frequency response of optical to electrical converters.
To create and use your own user-corrections, enter the User Correction menu.
(This menu is located under the 70820As Calib menu.) The microwave transi-
tion analyzer maintains two separate user-correction memories: one for chan-
nel 1 and one for channel 2. After entering the User Correction menu, the
microwave transition analyzer displays user-correction data in the lower half
of the screen. The screens upper half remains dedicated to showing the nor-
mal display. Or, if the displayed trace is assigned to the data, the trace shows a
graphical representation of the filter, as the following figure shows. To learn
how to assign a trace to user-corrections, refer to Viewing User-Correction
Data on page 3-56.