Eye-Diagram Analyzer Reference
Agilent 70820A Calibration
Viewing the Calibration Data
The ability to view the IF calibration data is mainly intended for service pur-
poses. However, the procedures in this section show you how to view the data.
To View IF
Correction Data
Traces
1 Display the Traces menu by pressing:
MENU, Traces
2 Turn on both trace 1 and trace 2. Set the input to trace 1 as channel 1. Set the
input to trace 2 to channel 2.
3 Display the IF correction data for each channel by pressing:
page 1 of 2, Calib, IF calib, cal debug, show ,
SHOW COR FFT or SHOW COR FIR
4 Scale the traces by pressing:
page 2 of 2, Scale
5 Select trace 1, and press:
AUTOSCALE
6 Select trace 2, and press:
AUTOSCALE