Getting Started
Steps for Setting Up Eye-Diagram Analysis
3 Turn the display off and select trace one by pressing:
Traces, display ON|OFF OFF
, select:, TR1
Move the
Measurement
Plane
To move the measurement plane, press:
Trg,Cal, CH1 EXT ATTEN and enter the attenuation on channel 1 input.
Optional Step. Save Instrument State as Preset State
1 Select the Traces menu and scale the screen to the displayed signal by
pressing:
Traces, AUTOSCALE
2 Save the current instrument state by pressing:
Trg,Cal, more 1 of 2, DEFINE USER IP, CONT
The current instrument state is saved, including the pattern generators HP-
MSIB address, the frequency reference, scaling, and trigger level. Whenever
INSTR PRESET is pressed, the eye-diagram analyzer is automatically placed
in these settings.