Module 7
Measurement Functions
7-31
Vth Measurement
Function Setup
Here are definitions for the Y1-axis plot (SQRT_ID) and the Y2-axis plot (PEAK).
@L1X means the X-axis intercept of LINE 1.
@L1G^2 means the squared gradient value of LINE 1. In this case, it is just calculating line 1 slope.
It is squared to account for the fact that the Y1-axis is a square root.
Click Help > Agilent EasyEXPERT Help to open the online help. And visit the Read Out Functions