Module 7
Measurement Functions
7-34
V/I DAC
Filter OFF
Filter ON
Time
Filter setting
SMU
(V or I mode)
Force
Sense
Series Resistor
(1 M ohm)
SMU Filter
The filter is mounted on each SMU. It assures clean source output with no spikes or overshooting.
However, using a filter may increase the SMU settling time. If measurement speed is top priority, set
the SMU Filter OFF.