Agilent Technologies EasyEXPERT Guida Utente Pagina 165

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 344
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 164
Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 2- 39
Classic Test Definition
Direct Control
compenImZ
This built-in function performs compensation on the specified impedance data and
returns the imaginary data for the compensation results.
This function is effective for the CMU measurement data compensation while the
switching matrix is used with the CMU to extend the measurement terminals.
Define the compenReZ and compenImZ functions as the user functions on the
Function Setup screen, so that the compensation is performed while the Direct
Control test is being executed. Also enter the formula to calculate the measurement
parameters. An example definition is shown in
Figure 2-2. For the equations, see
“About Impedance Measurements” on page 4-22.
Syntax compenImZ(frequency,Real,Imaginary)
Parameters frequency Measurement frequency. Hz. Numeric scalar or vector data.
For the C-f measurement data, frequency should be a vector
variable that stores multiple frequency values.
Real CMU Re(Z) measurement result data. . Numeric vector data.
Real should be a vector variable that stores the impedance real
data before compensation.
Imaginary CMU Im(Z) measurement result data. . Numeric vector data.
Imaginary should be a vector variable that stores the impedance
imaginary data before compensation.
Example Name Xcompen
Definition compenImZ(Freq,ReZ,ImZ)
Xcompen returns the impedance imaginary data for the compensation results.
Vedere la pagina 164
1 2 ... 160 161 162 163 164 165 166 167 168 169 170 ... 343 344

Commenti su questo manuale

Nessun commento