Agilent Technologies EasyEXPERT Guida Utente Pagina 280

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 344
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 279
4- 36 Agilent EasyEXPERT Users Guide Vol. 1, Edition 1
Function Details
SPGU Module
In the example in Figure 4-19, SPGU1 and SPGU2 are used. The channels output
Pattern2 twice and Pattern1 three times in one sequence, although the two patterns
for SPGU1 are the same.
The SPGU channels simultaneously start the sequence output. The sequence will be
repeated in the same way as for the pulse period. The channels continue the
sequence output until the specified time elapses or the repeat count reaches the
specified number of sequences.
NOTE
Setup delay time between the different patterns
When the pattern is changed to a pattern of a differing index, a 50 ns setup delay will
always be inserted.
There is no delay between patterns of the same index.
Vedere la pagina 279
1 2 ... 275 276 277 278 279 280 281 282 283 284 285 ... 343 344

Commenti su questo manuale

Nessun commento