Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 4- 7
Function Details
I/V Sweep Measurement
2. Set the VAR1 channel as shown in “Basic Sweep Measurement”.
3. Specify the following parameters for VAR2 on the Measurement Setup tab
screen.
Start Start value of secondary sweep. Allowable range of start
depends on the output range of secondary sweep source.
Stop Stop value of secondary sweep is calculated from start, step,
and number of steps. Allowable range of stop depends on
the output range of secondary sweep source.
Step Step increment of secondary sweep.
No of Step Number of secondary sweep steps (N
2
). 1 to 1001.
Then, the following formula must be satisfied.
1 N
1
N
2
128128 (N
1
: number of primary sweep steps)
Compliance Compliance value of secondary sweep source. Allowable
range of compliance depends on the compliance range of
secondary sweep source.
Pwr Comp (Optional) Power compliance value of secondary sweep
source. Allowable range of power compliance depends on
the power compliance range of sweep source.
Figure 4-4 Subordinate Sweep Measurement