Agilent Technologies HP8648A Manuale Utente Pagina 111

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 320
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 110
FM Receivers
110
On the Test Set using the knob and data entry keys:
14 Change SINAD meter to read Distortion.
Distortion is displayed as shown.
15 Select AC Level.
16 Press the REF SET key.
17 Decrease the Receiver’s volume control until the Test Set’s AC Level
meter reads 17 dB.
Distortion displayed as shown above.
Distortion
To Screen
RX TEST
AC Level
Filter 1
50Hz HPF
Filter 2
15kHz HPF
Atten Hold
On/Off
Ext Load R
8.00
RF GEN
RF ANL
AF ANL
SCOPE
SPEC ANL
ENCODER
DECODER
RADIO INT
More
dB
%
-17.18
Distn
RF Gen Freq
145.280000
MHz
Amplitude
-47.0
dBm
Output Port
RF Out/Dupl
AFGen1 Freq
1.0000
kHz
AFGen2 Freq
1.0000
kHz
AFGen2 To
Off
FM
Ref
AFGen1 To
3.00
kHz
FM
1.2
Vedere la pagina 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 319 320