Agilent Technologies HP8648A Manuale Utente Pagina 121

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 320
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 120
AM Transmitters
120
Measurement Procedure:
1 Connect the Antenna to the ANT IN connector.
CAUTION: Do not exceed the connector’s rated input or permanent instrument damage
may result.
On the Test Set:
2 Press the PRESET key.
Using the knob and data entry keys:
3 Set AF Anl In to AM Demod.
4 Select the SPEC ANL screen.
5 From the Controls select Ant.
6 Set Center Freq and Span fields to view desired spectrum.
7 Set Ref Level from 30 dBm to 50 dBm as required to view the desired
signal.
Test Set Options Required
Spectrum Analyzer/Tracking Generator (option 102)
Additional Equipment
Required
Whip antenna
Vedere la pagina 120
1 2 ... 116 117 118 119 120 121 122 123 124 125 126 ... 319 320

Commenti su questo manuale

Nessun commento