Using the Tracking Generator
224
17. Select Main from the Choices field.
Pass Cavity:
18. Connect the pass cavity as shown.
19. Adjust the tuning rod to the desired pass frequency.
20. Adjust the coupling loops for desired band-pass/insertion loss.
NOTE: If Sensitivity dB/div is changed from 10 dB/div the coaxial cables must be
shorten, and the previous steps performed on the Auxiliary screen must be
repeated to re-normalize the display for the new setting.