On the Test Set using the knob and data entry keys:
14. Decrease Amplitude until the SINAD meter reads 10 dB.
If desired, use the meter averaging function for the SINAD indicator.
a. Select dB on the SINAD meter.
b. Press the AVG key.
If desired, select the number of readings to average.
Enter the number of readings to average.
Higher number of readings averaged will require longer to
reach the final indication.
15. Set Amplitude units to dBm.
16. Increase Amplitude (more positive) by 3 dB.
Record Amplitude setting for later use.