High Voltage Differential Active Probes
• 25 MHz to 800 MHz bandwidth
• Switchable attenuation
• Measure up to 1,400 V CAT II and
7 kV CAT I
N2790A/91A and N2891A high
voltage differential probes
Oscilloscope users often need to
make floating measurements where
neither point of the measurement
is at earth ground. Use N2790A,
N2791A or N2891A high voltage
differential probe to make safe and
accurate floating measurements
with oscilloscope. The N2790A,
N2791A and N2891A high voltage
differential probes allow conventional
earth-grounded Agilent oscilloscope
to be used for floating signal
measurements.
Each probe offers user selectable
attenuation settings that make it
highly versatile, allowing it to be used
for a broad range of applications.
The probe comes with probe tip
accessories for use with small and
large components in tight spaces.
The N2791A and N2891A are
compatible with any oscilloscope
with 1 Mohm BNC input. The
N2791A and N2891A probe power is
supplied by included 4x AA batteries
or USB host port of the scope or
PC via a supplied USB power cable.
The N2790A is compatible with the
Agilent’s AutoProbe interface where
the probe power is supplied by the
Agilent oscilloscope’s probe interface.
1141A 200 MHz differential
probe
The 1141A is a 1x FET differential
probe with 200-MHz bandwidth and
a 3000:1 CMRR. The probe has a
high-input resistance and low input
capacitance of 7-pF to minimize
circuit loading. The 1141A must be
used with 1142A probe control and
power module, which controls input
coupling modes dc, dc with variable
offset, and dc reject. Two attenuators,
10x and 100x are provided to expand
the linear differential input range to
±30V. This probe works with any 50-Ω
input oscilloscope including 5000,
6000 and 7000 Series.
1141A 200 MHz differential probe with
1142A probe control and power module
Characteristics for N2790A, N2791A and N2891A differential probe
N2790A N2791A N2891A
Bandwidth 100 MHz 25 MHz 70 MHz
Risetime 3.5 nsec 14 nsec 5 nsec
Attenuation ratio 50:1 / 500:1 10:1 / 100:1 100:1 / 1000:1
CMRR -80 dB at 50/60 Hz
-50 dB at 1 kHz
-50 dB at 1 MHz
-80 dB at 50/60 Hz
-40 dB at 1 MHz
-80 dB at 50/60 Hz
-60 dB at 20 kHz
Max input voltage
to ground
±1000 V (CAT II)
±600 V (CAT III)
±700 V at 100:1
±70 V at 10:1
±7000 V at 1000:1
±700 V at 100:1
Max input voltage
between two
inputs
±1400 V at 500:1
±140 V at 50:1
±700 V at 100:1
±70 V at 10:1
±7000 V at 1000:1
±700 V at 100:1
Max number
of probes
supported by
5000/6000/7000
Series
444
Characteristics for 1141A differential
probe
1141A
Bandwidth 200 MHz
Risetime 1.75 nsec
Attenuation ratio 10:1 and 100:1
with attenuater
High CMRR 3000:1 at 1-MHz
10:1 at 100-MHz
Input
impedance
Between inputs:
1 MΩ, 7 pF
Max input voltage 200 Vdc + peak ac
(probe alone)
500 Vdc + peak ac
(with attenuator)
Max number
of probes
supported by
5000/6000/7000
Series
4