11-20
Agilent Technologies E5500 Phase Noise Measurement System
FM Discriminator Measurement Examples
Discriminator Measurement using FM Rate and Deviation Calibration
Defining the
Measurement
1. From the File menu, choose Open.
2. If necessary, choose the drive or directory where the file you want is
stored.
3. In the File Name box, choose “vco_r&d.pnm”.
4. Click the Open button.
The appropriate measurement definition parameters for this example
have been pre-stored in this file. Table 11-3 on page 11-16 lists the
parameter data that has been entered for the FM discriminator
measurement example.)
Vedere la pagina 308