Agilent Technologies E5503A Guida Utente Pagina 317

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 465
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 316
11-28
Agilent Technologies E5500 Phase Noise Measurement System
FM Discriminator Measurement Examples
Discriminator Measurement using FM Rate and Deviation Calibration
Making the
Measurement
1. press the Continue key when you are ready to make the measurement.
Calibrating the Measurement
The calibration procedure determines the discriminator constant to use in the
transfer response by measuring the system response to a known FM signal.
NOTE Note that the system must be operating in quadrature during calibration.
2. First establish quadrature by adjusting the phase shifter until the meter
indicates 0 volts, then press Continue.
Vedere la pagina 316
1 2 ... 312 313 314 315 316 317 318 319 320 321 322 ... 464 465

Commenti su questo manuale

Nessun commento