14-6
Agilent Technologies E5500 Phase Noise Measurement System
Baseband Noise Measurement Examples
Baseband Noise without using a Test Set Measurement Example
Baseband Noise without using a Test Set Measurement
Example
This measurement example will help you measure the noise voltage of a
source.
NOTE To ensure accurate measurements, you should allow the UUT and
measurement equipment to warm up at least one hour before making the
noise measurement.
Defining the
Measurement
1. From the File menu, choose Open.
2. If necessary, choose the drive or directory where the file you want is
stored.
3. In the File Name box, choose “BBnoise_without_testset_89410.pnm”.
Vedere la pagina 355