Agilent Technologies AGILENT U2761A Manuale di Servizio Pagina 116

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 169
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 115
10 Arbitrary Waveform Commands
108 U2761A Programmer’s Reference
DATA:ATTRibute:CFACtor?
Syntax
DATA:ATTRibute:CFACtor?
This query returns the crest factor of all data points for the Arbitrary
waveform.
Remarks
Crest factor is the ratio of a signal peak value to its RMS value and will
differ according to waveshape. The crest factor can be computed as
follows:
Crest factor = Vpeak / Vrms
The typical crest factor values for several common waveforms are stated
in the following table:
Example
The following query returns the crest factor of the data points. (Assuming
that the Arbitrary waveform used has a sine waveshape).
DATA:ATTR:CFAC?
Typical response:
+1.4142140000000E+00
Waveform Crest factor
Sine 1.414
Triangle 1.732
Square 1
DC 1
DATA :ATTR :CFAC ?
Vedere la pagina 115