14 CALibration Commands
130 U2761A Programmer’s Reference
CALibration:SECure:STATe
Syntax
CALibration:SECure:STATe {OFF|ON|RESET}, <code>
This command unsecures or secures the U2761A for calibration. If you
forget your security code, you can reset the code by specifying U2761A.
CALibration:SECure:STATe?
This query returns the security state as 0 (OFF) if the U2761A is
unsecured, or 1 (ON) if secured.
Parameters
Examples
The following command secures the U2761A for calibration.
CAL:SEC:STAT ON, U2761A
The following query returns the security state.
CAL:SEC:STAT?
Typical response:
1
Item Type Range of values Default value
state Discrete OFF|ON|RESET ON
code Numeric Up to 12 alphanumeric characters U2761A
CAL space:SEC OFF
ON
RESET
:STAT , code