APPLy Commands 3
U2761A Programmer’s Reference 33
APPLy:SINusoid
Syntax
APPLy:SINusoid [<frequency> [,<amplitude> [,<offset>] ]]
This command outputs a Sine wave with the specified frequency,
amplitude, and DC offset. The waveform is output as soon as the
command is executed.
Parameters
Example
The following command sets the output frequency to 2 kHz, output
amplitude to 1 Vpp, and output offset to 0 V.
APPL:SIN 2000, 1 VPP, 0
Item Type Range of values Default value
frequency Numeric 1 μHz to 20 MHz 1 kHz
amplitude Numeric 40 mVpp to 5 Vpp (Into 50 Ω load)
80 mVpp to 10 Vpp (Into open circuit)
1 Vpp
offset Numeric ±2.48 V (Into 50 Ω load)
±4.96 V (Into open circuit)
0 V
APPL :SIN space frequency , amplitude , offset