4 Output Configuration Commands
52 U2761A Programmer’s Reference
[SOURce:]FUNCtion:RAMP:SYMMetry
Syntax
[SOURce:]FUNCtion:RAMP:SYMMetry <percent>
This command sets the symmetry percentage for the Ramp wave.
[SOURce:]FUNCtion:RAMP:SYMMetry?
This query returns the current symmetry setting in %.
Parameter
Examples
The following command sets the symmetry to 50%.
FUNC:RAMP:SYMM 50
The following query returns the symmetry.
FUNC:RAMP:SYMM?
Typical response:
+5.0000000000000E+01
Item Type Range of values Default value
percent Numeric 0% to 100% 100%
SOUR : FUNC space
?
:RAMP :SYMM percent
NOTE
Setting the symmetry to 50% will produce a Triangle wave, and 0% for a Negative Ramp
wave.