Phase-Lock Commands 12
U2761A Programmer’s Reference 117
[SOURce:]PHASe[:ADJust]
Syntax
[SOURce:]PHASe[:ADJust] <angle>
This command adjusts the phase offset of the output waveform in degrees
or radians as specified by the UNIT:ANGLe command.
[SOURce:]PHASe[:ADJust]?
This query returns the adjusted phase offset in the specified unit.
Parameter
Examples
The following command sets the phase offset to 180 °.
PHAS 180
The following query returns the phase offset.
PHAS?
Typical response:
+1.8000000000000E+02
Item Type Range of values Default value
angle Numeric –360 ° to 360 ° (for degrees)
–2 p to 2 p (for radians)
0 °
SOUR : PHAS space
?
angle:ADJ