14 CALibration Commands
132 U2761A Programmer’s Reference
CALibration:SETup
Syntax
CALibration:SETup <1|2|3| . . . |115>
This command configures the U2761A internal state for each of the
calibration steps to be performed. Before any calibration step is performed
for the U2761A, the default setup is 0.
CALibration:SETup?
This query reads the calibration setup number and returns a value from 1
to 115.
Parameter
Examples
The following command sets the U2761A calibration setup number to 5.
CAL:SET 5
The following query returns the calibration setup number.
CAL:SET?
Typical response:
+5
Item Type Range of values Default value
setup number Numeric 1 to 115 0
spaceCAL 1
2
3
...
115
:SET
?
Vedere la pagina 139